Touchclave Ecotech
24
2.4 Typical cycles
Though the Touchclave Ecotech is a fully programmable autoclave, the table below shows details of
some typical operating cycles, showing set values for different types of load.
Note: Cycle development may be required to optimise cycle performance and effectiveness. Contact
LTE Service Centre for more information on cycle development.
Standard Cycles
OPERATING CYCLE
NAME
CYCLE OR
LOAD TYPE
STERILIZING
TEMPERATURE
STERILIZING
TIME
MEDIA
Fluid
121
o
C
15 minutes
FLUID DISCARD
Fluid
134
o
C
5 minutes
PLASTIC DISCARD
Plastic
Discard
134
o
C
5 minutes
EMPTY GLASSWARE
Empty
Glassware
134
o
C
5 minutes
MELT/FREE STEAM
Melt/
Free Steam
105
o
C
30 minutes
INSTRUMENT
Instruments 134
o
C
5 minutes
2.5 Technical Specification
Electrical
Electrical supply voltage (RMS)/ Frequency: 240V AC/ 50Hz
Electrical supply current: 12.5A (Maximum)
Environmental
Altitude: 2000m (Maximum)
Ambient temperature: 5 -
40⁰C
Ambient Relative Humidity: 20
–
80% (non-condensing)