Se�ngs 2/3
Se�ngs 3/3
(32)
Either specifies the period for automa�c dimming of
the colour display or deac�vates the func�on:
30 seconds, 1 minute, 5 minutes, off.
(33)
Either specifies the period for automa�c switch-off of
the colour display or deac�vates the func�on:
10 minutes, 30 minutes, 1 hour, off.
NOTE: This func�on is deac�vated during an automa�c
measurement.
(34)
Switches signal tones/key tones on or off.
(35)
Opens the next screen.
(36)
Configures the duty factor at the sensor:
normal:
duty factor 50%
sqrt(2):
duty factor 33%
Changing the duty factor enables the determina�on of
the sensor‘s self-hea�ng. This measuring method can
be used for high-precision measurements at fixed
points, such as the water triple point, and in suitable
media ensuring a high temperature stability.
The selected se�ng is displayed on the screen if it
differs from the se�ng „Normal“.
At this model, the lineariza�on fuc�on is disabled, as
the supplied sensor data is specially integrated.
At this model, no coefficients can be entered, as the
supplied sensor data is specially integrated.