4.5 Measure Continuity
Cable run test,
potential free contact.
<100.0Ω is closed
5.0 Measuring
Select
in the main menu then press the
key.
„Measure“
ENTER
5.1 Milliamps
Select
then press the
key for the milliamp measuring submenu.
„Milliamps“
ENTER
Range
(LEFT
RIGHT
or
keys to change)
0 - 20 mA
,
or
.
4 - 20 mA
Function
(LEFT
RIGHT
or
key to change)
the displayed % is linear to the mA input,
Linear:
or
the displayed % is the square root of
Flow:
the mA input.
Main Unit
(LEFT
RIGHT
or
key to change)
mA:
the primary display shows the mA,
or
%:
the primary display shows the %.
Loop Power
(LEFT
RIGHT
or
keys to change)
Off:
the loop is externally-powered
or
On:
the unit supplies 24V to power the loop.