Ramp procedure effecve example:
Let‘s say, that the set temperature is the ambient one and that it is necessary to reach 400°C with a
gradient of 2°C/min.
• Press „F“-key, set „SP2" to „400°C“ using the arrow-keys. Press „E“-key to confirm the choice.
• Press „F“-key, set „GRAD“ to „2°C/min“ using the arrow-keys. Press „E“-key to confirm.
• Press „F“-key, set „RAMP“ to „ON“ using the arrow-keys. Press „E“-key to confirm.
Aer pressing the „E“-key for confirming the ramp start, the calibrator will acsend with the set
slope. Of course, there will be some oscillaons at the beginning since the ramp slope will not be
suitable but they will stop in a short me and then the calibrator temperature will follow the
ramp‘s set point one.
•
RES
Display reading resoluon (0.1/0.01): the values admi ed are „0.1" and „0.01" and they can be
selected by the arrow-keys.
•
SW/ON
Thermostat Switch ON: it displays the temperature at which the thermostat‘s contact, connected
to the „SWITCH TEST“ bushes, is closed.
•
SW/OFF
Thermostat Switch OFF: it displays the temperature at which the thermostat‘s contact, connected
to the „SWITCH TEST“ bushes, is open.
The value is reset each me the power supply fails or by pressing both arrow-keys at the same
me. The value is updated every me that the thermostat‘s contact closing is detected.
•
- only for
versions with measuring inputs
SENSOR
-2I
This parameter allows at
versions of the
temperature calibrators (with 2 measuring
-2I
LR-
Cal
inputs for external reference and one test item or for two test items) the reading of connected
sensors.
OFF:
no input is enabled to read the sensor‘s value.
EXT:
the 4 bushes of the input „EXT“ are enabled to read the sensor connected to them,
whose value is indicated at the 2nd. line of the display
REF:
the 4 bushes of the input „REF“ are enabled to read the sensor connected to them,
whose value is indicated at the 2nd. line of the display. (except model
)
BK40-M-2I
LR-
Cal
EXT+REF: the 8 bushes of the 2 inputs „EXT“ and „REF“ are enabled to read the sensors
connected to them, whos values are indicated at the 2nd. line of the display.
(except model
).
BK40-M-2I
LR-
Cal