4. Installaon and Operaon
4.1 Installaon
4.1.1 Removal of packaging
The temperature calibrator is equipped with suitable packaging for transport and tradional
shipping systems. Any damage caused during transport must be nofied immediately to the
transporters and claim must be made.
4.1.2 Posioning the temperature calibrator
Posion the temperature calibrator on a safe and clean place.
The temperature calibrator is suitable for operang at high temperatures with the
consequent danger of fire. Keep it away from any type of inflammable materials
and never put any type of liquid inside (except the liquid reservoir).
To avoid any smell in the room it is be er to switch on the temperature calibrator
outside the room or to use a fume hood to remove any vapours given off by hot
bath fluid.
4.1.3 Power supply
The temperature calibrator runs on a voltage of 230 VAC (115 VAC on request), single-phase,
50/60 Hz. A 2.5 meter long cable 3 x 1 mm² is supplied with the instrument.
Make sure that the plant is earthen correctly before switching on the temperature calibrator.