8.7 Fuses, commuting the device voltage
CAUTION! Before working on the fuses or opening the device, make sure
that the power cord is removed!
In the LPKF ProtoMat 93s, the primary and secondary voltages are fused. The
main fuses (primary) are in the power connection of the control unit and are
accessible from the outside.
Caution! When switching to another line voltage make sure that
both
fuses are
exchanged. Both fuses must be of the
same
value.
1,6 A
3,15 A
230 V
115 V
Voltage setting and fuses
The secondary fuses are inside the control unit. The control unit must first be
removed from the mechanical part to replace the fuses. To do this, loosen screws
1, 2, 3 and 4 and the M4 screws 5 and 6 at the bottom of the device and then
remove the cable connections between the mechanical part and the control unit.
1
2
3
4
5
6
Position of screws for removing the electronics from the mechanical part of an LPKF ProtoMat 93s
CAUTION! Be sure the power cable is removed!