31
Operating data
{
{
Warm-up time:
max. 30min., regularly 10-15min
{
{
Recommended
flow rate:
5 to 60l/min
{
{
Humidifier
system output:
> 10mg/l in range 5 - 60l/min
{
{
Maximum
operating pressure:
200mbar
(1)
{
{
Continuous noise:
< 50dB (1m)
{
{
Decibel level
of alarms:
max. 65dB
{
{
Max. volume of water: 200 ml
humidification
system
{
{
Gas leakage at max.
operating pressure:
< 10 ml/min
(2)
{
{
Gas leakage
at 60mbar:
< 5 ml/min
(2)
{
{
Pressure drop:
< 0.02 (mbar*min)/l
(2)
{
{
Internal Compliance:
Minimum 1.0 ml/mbar
(2)
Maximum 2.0 ml/mbar
(2)