6 Function check
6.1 Intervals
Carry out a function check at regular intervals:
• After each hygienic preparation
• After each repair
• At least every 6 months
6.2 Performing a function check
1. Inspect the housing for cracks, damage, and heavy soiling.
2. Connect the module up to the therapy device
Connecting the module up to the therapy device", page
3. Check whether the
symbol is shown on the therapy
device display.
4. Connect the pulse oximetry sensor
5. Check whether the current oxygen saturation and pulse rate
values are shown on the display.
6. If the module does not work properly or shows signs of
damage: Contact your authorized dealer.
Result
The function check is complete.