Lowenstein Medical MiniScreen plus Description And Instructions For Use Download Page 5

2.  Information

2.1.  Safety-related informations

 

 Observe the instruction manual

:

Every use of the device requires exact knowledge
and observance of this instruction manual. The
device is only intended for the use described herein.

  

 

 No alarm function available!

The device is not suitable for continuous monitoring
of vital or physiological functions (for example,
intensive monitoring, monitoring operation) since
there is no SpO2 alarm. No direct data analysis takes
place in the device.

 

 No SIDS Monitoring

:

The device is not suitable for use as a SIDS monitor
(SIDS: Sudden Infants Death Syndrome, sudden
child death).

  Patient instruction

:

Patient instruction must be provided by a healthcare
professional trained in the device. An attached short
manual does not replace the manual instructions or
the warning of possible hazards.

   

  Separation from the supply network:

:

To isolate the device from the supply network, the
plug of the power supply must be disconnected.

  

Do not open the device!

   

  

Warning:

Additional devices that are connected to medical
electrical (ME) devices must be able to meet their
IEC or ISO standards. Anyone connecting additional
devices to ME devices is a system configurator and
is responsible for the ME system being in compliance
with the normative requirements (for example, IEC
60601-1).

  

Warning:

For online measuring you must use an electrical or
optical isolator for connection of the device to the PC
(available from your home care provider). A connec-
tion of the device to the PC without electrical or
optical isolator is only allowed if all patient connec-
tions have been removed first. Only Physicians or
trained staff is allowed to perform online measure-
ments.

  

  

Warning:

Magnetic and electrical fields can impact on the func-
tioning of the device. Use of this equipment adjacent
to or stacked with other equipment should be
avoided because it could result in improper ope-
ration. If such use is necessary, this equipment and
the other equipment should be observed to verify that
they are operating normally. When operating the
device, ensure that all third-party devices being

operated in the vicinity are meeting their relevant
EMC requirements. X-Ray equipment, HF surgical
devices, tomography devices etc. can interfere with
other devices as they may emit higher levels of
electromagnetic interference.

  

  

Warnung:

The device has no defibrillator protected application
parts! Disconnect the device before defibrillation! A
direct application of the device to the heart (esp.
ECG!) is not allowed!
While the device is being worn, no invasive or
intracorporal measurements and interventions (e.g.,
electro- or HF-surgical devices) may take place.

  

  

Warning:

Using multiple devices for patient screening could re-
sult in excessive leakage current above the allowed
value!

  

  

Warning:

When using the device on patients with active
implants, such as Cardiac pacemakers, peripheral
nerve stimulation devices, tongue pacemaker etc.,
medical professionals should be alert to possible
interference with the device or the implant.

  

  

Warning:

Applying the device to more than one patient at any
one time is not allowed.

  

Warning:

The device must not be operated in vehicles and
aircraft.

  

Warning

:

When applying sensors to the patient, please ensure
that no electrical line of the device is in contact with
any

other

electrical

conductive

parts

including

ground.

  

  

Warning

:

Avoid placing this instrument in direct sunlight or in
close proximity to intense heat. Prevent also the
contact with dust, lint, dirt, moisture and liquids. 

  

  

Warning

:

Children or incompetent persons should not use this
device unattended without first having obtained ins-
tructions for the safe use of the device. Ensure that
infants, children and animals cannot touch the device
unattended

- 3 -

Summary of Contents for MiniScreen plus

Page 1: ......

Page 2: ......

Page 3: ...and abdominal movement 5 3 1 6 Sensor for body position 6 3 1 7 Sensor for leg movement 6 3 1 8 Sensor for ECG 6 3 1 9 Sensor for Neuroport 6 3 2 Preparing for measurement ambulatory 6 3 3 Operating...

Page 4: ...s incl connection cable It is also equipped with the following applied parts SpO2 finger sensor with cable for obtaining pulse frequency and oxygen saturation values Velcro strap for securing the fing...

Page 5: ...of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper ope ration If such use is necessary this equipment and the other equipment should be...

Page 6: ...ing the end of its service life the device and its accessories should be disposed of in accordance with the WEEE Directives or relevant Electronic Disposal protocol For further details please contact...

Page 7: ...turation and pulse frequency A failure of the finger sensor or a missing data update by the pulse oximeter is indicated to the user by the red LED on the device At the same time the values for SpO2 an...

Page 8: ...be cleaned with a sterile solution A headband around the forehead will eliminate movement of the electrode during the study The blue marked connecting plug of the EEG electrode cable needs to be plug...

Page 9: ...sensor If applicable attach the abdominal belt leg sensor ECG and EEG electrodes Check that the sensors and their connectors are positioned correctly The next morning The patient has to detach the Min...

Page 10: ...ominal straps can be washed in the washing machine at 60 C use a laundry net Let the straps dry in the air Attention Do not use a dryer Remove thoracic and abdominal sensor beforehand Disinfection It...

Page 11: ...tion of the PAP adapter hose on the oxygen mask and on the MiniScreen unit Signal amplitudes are very small or do not appear Check the corresponding sensors on the patient and their connections to the...

Page 12: ...detection of leg movements restless legs 200 0936 USB cable for data transmission between MiniScreen and PC 100 0815 03 Electrical isolator USB for galvanic separation between MiniScreen device and PC...

Page 13: ...ensor special rubber coated thimble finger sensor Pulse wave Plethysmogramm display measurement via finger sensor Position Acceleration sensor for position recording 5 positions Ambient Light Photomet...

Page 14: ...9 Used Symbols 12...

Page 15: ...SPR 11 Class B The MiniScreen is suitable for the use in all establish ments including domestic establishments and those which are connected to the public power supply which powers buildings for resid...

Page 16: ...ateur bands 80 AM 1 kHz Radiated RF disturbance IEC 61000 4 3 Proximity fields from RF wireless communications equipment IEC 61000 4 3 10 V m 80MHz 2 7 GHz 80 AM 1 kHz 380 390 MHz 27 V m PM 50 18 Hz 4...

Page 17: ...0 3 28 5 240 5 100 5 800 WLAN 802 11 a n Pulse Modulation b 217 Hz 0 2 0 3 9 5 500 5 785 NOTE If necessary to achieve immunity test levels the distance between the transmit antenna and the MiniScreen...

Page 18: ...ance 8 Neuroport 6 Operating elements 6 Optical isolator 3 Ordering informations 10 PAP 2 PAP sensor 5 Position sensor 6 Preparing for measurement 6 Pressure pads 5 RLS sensor 6 Safety related informa...

Page 19: ......

Page 20: ...18...

Reviews: