4 Preparation and operation
12
|
EN
LM
T
68
54
1a
0
6/
20
21
4.2.1 Connecting leakage circuit
1. Push the circuit onto the device outlet port.
2. Connect the invasive or non-invasive patient/ventilator
interface to the leakage circuit (see instructions for use
for the patient/ventilator interface).
4.2.2 Connecting single circuit with
valve
1. Push the free end of the circuit onto the device outlet
port.
2. Connect the pressure measuring tube to connection
.
3. Connect the valve control tube to connection
.
4.2.3 Connecting double circuit
1. Push the free end of the circuit
port.
2. Push circuit
onto the device inlet for exhaled air.
3. Connect the pressure measuring tube
to connection
.
4. Connect the patient/ventilator interface (e.g. mask) to
the Y-piece of the circuit.
4.2.4 Connecting circuit for mouthpiece
ventilation
1. Push the free end of the circuit onto the device outlet
port.
2. Connect mouthpiece to the circuit (see instructions for
use for the patient/ventilator interface).
Risk of injury if patient valve is covered!
If the patient valve is covered, exhaled air can no longer be
routed away and the patient will be put at risk.
Always keep the patient valve free.
P
As an alternative to the leakage circuit, it is also
possible to use a single circuit with valve or double
circuit for mouthpiece ventilation.