7 Alarms
26
|
EN
LM
T
68
65
1c
0
2/
20
22
Error internal battery
communication
572
Internal battery defective.
Device defective.
Contact your specialist dealer.
Error battery E1
communication
573
External battery 1 defective.
Device defective.
Contact your specialist dealer.
Error battery E2
communication
574
External battery 2 defective.
Device defective.
Contact your specialist dealer.
Error battery E1
575
External battery 1 defective.
Contact your specialist dealer.
Error battery E2
576
External battery 2 defective.
Contact your specialist dealer.
Error internal battery
temperature
577
Ambient temperature too high.
Operate device at an ambient temperature of
5 °C to 40 °C.
Error battery E1
temperature
578
Ambient temperature too high.
Operate device at an ambient temperature of
5 °C to 40 °C.
Error battery E2
temperature
579
Ambient temperature too high.
Operate device at an ambient temperature of
5 °C to 40 °C.
Power outage
580
Power supply failed.
Use alternative ventilation option.
Energy supply via
internal battery
584
Power supply failed.
Check that the power cord is securely
connected. Check function of socket.
External battery and power supply not
connected.
. If
necessary, connect power supply.
No exhalation valve
753
No exhalation system present.
Check circuit and patient interface.
Connect exhalation system.
Pressure permanently
low
755
Mask leakage too high.
Check and correct position of mask.
Tidal volume
permanently low
756
Settings implausible.
Check therapy and alarm settings.
Exhalation blocked
757
Exhaled air outlet is blocked.
Check exhalation system and expiration
module.
Constant pressure level
758
Respiratory frequency or set pressure
difference too low.
Check therapy settings.
Intake area blocked
759
Intake area blocked.
Keep intake area free.
Pressure measuring and
valve control tubes
switched
760
Valve control and pressure measuring
tubes switched.
Check circuit for correct fit
.
Valve control tube kinked.
Check valve control tube for blockages and
damage.
If necessary: Replace circuit.
DISPLAY
CODE
CAUSE
ACTION