5 Settings in the menu
LM
T 6
86
51
c 0
2/2
02
2
EN
|
17
5.2.2 Report menu in the Patient menu
(usage data)
Information about the parameters in this menu can be found
in the table below.
5.2.3 System menu in the Patient menu
5.2.4 Device settings submenu
PARAMETER
DESCRIPTION
Alarm list
Lists the alarms which have occurred.
The log is retained when the alarm
system or the device is switched off.
The start and end of ventilation is
recorded.
The log is retained even if the device is
disconnected from the power supply
and the batteries are removed.
The log can store 1,000 alarms. Once
this capacity limit has been reached,
the oldest alarm is deleted and the
new alarm is saved.
Event list
Lists the events that have occurred.
Alarm/event list
Lists the alarms and events which have
occurred in chronological order.
Parameter overview
Lists all parameters and set values for
the up to 4 ventilation programs
which can be configured.
Device usage
Obtain information here about the
patient’s therapy (duration, days used,
program proportions) and about
device usage (operating time, internal
battery life remaining or charging
state of internal battery in percent).
Circuit test
Perform a circuit test here on change
of patient and as required. This checks
for resistance, compliance, and leaks
(see “4.5 Performing circuit test”,
page 14)
FiO
2
cell
Activate or deactivate the FiO
2
cell
and calibrate the FiO
2
cell here.
Export therapy data
You can export the set device settings
here. A USB flash drive must be
connected for exporting.
Flight mode
You can activate/deactivate flight
mode here. With flight mode
activated, all wireless communication
(Bluetooth) is ended.
Device settings
You can configure the device here
(see “5.2.4 Device settings
submenu”, page 17)
Device status
Obtain information here about the
device (name, type, serial number of
device and components, firmware
version) and about the internal
battery.
PARAMETER
DESCRIPTION
Alarm volume
The patient can set the alarm level here.
1= very quiet, 2= quiet, 3= loud,
4= very loud
You can test the alarms here.
Display
You can set brightness and display
background here.
Filter timer
You can activate and reset the filter
change reminder function here.
Date and time
You can set the current date and time
here.
Connectivity
You can activate the Bluetooth
function and pair the device with the
LUISA app here.