Operation | 7
Basic Instructions for Use elisa 300-500
|
SW 2.09.x
|
Rev. 11
91
7.3.7
Overview of alarm limits
7.3.7.1
Ventilation alarms
MV max
Upper alarm limit for the expiratory minute volume
MV min
Lower alarm limit for the expiratory minute volume
Vt max
Upper alarm limit for the expiratory tidal volume
Vt min
Lower alarm limit for the expiratory tidal volume
Leakage
Criterion for the “Leakage high” alarm. The percentage describes the
leakage rate which is calculated from the total values (100 x (TVi-TVe)/
TVi).
Plimit
(relative or abso-
lute)
Maximum peak airway pressure
P min
Minimum pressure within a breath
PEEP max
Upper alarm limit for the measured PEEP
PEEP min
Lower alarm limit for the measured PEEP
RR max
Upper alarm limit for the measured respiratory rate
RR min
Lower alarm limit for the measured respiratory rate
Remark
Depending on the selected configuration, the alarm limit for the
maximum peak airway pressure PLimit can be set manually or it is
linked with the Pinsp setting and adjusted automatically.
7.3.7.2
Gas
+ O
2
max
Upper alarm limit for the inspiratory oxygen concentration. The dif
-
ference between the alarm limit and set inspiratory oxygen concen
-
tration (O
2
) is maintained if the O
2
value changes. The alarm criterion
can be switched off at the upper end of the adjustment range. In this
case, the “O
2
limit values?” message appears.
– O
2
min
Lower alarm limit for the inspiratory oxygen concentration. The dif
-
ference between the alarm limit and set inspiratory oxygen concen
-
tration (O
2
) is maintained if the O
2
value changes. The alarm criterion
can be switched off at the lower end of the adjustment range. In this
case, the “O
2
limit values?” message appears.
etCO
2
max
Upper alarm limit for the expiratory CO
2
concentration
etCO
2
min
Lower alarm limit for the expiratory CO
2
concentration
inCO
2
max
Upper alarm limit for the inspiratory CO
2
concentration
inISO max
Upper alarm limit for the inspiratory isoflurane concentration
inISO min
Lower alarm limit for the inspiratory isoflurane concentration
Summary of Contents for elisa 300
Page 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Page 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Page 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Page 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Page 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Page 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Page 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Page 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...