The manoeuvre ends when the set PEEP is reached (see Set ven
-
tilation parameters).
A 2.6.5
Performing the PEEPfinder
manoeuvre (inspiratory and
expiratory)
The ventilation alarms are disabled during the manoeuvre, and the ventilation
mode window indicates “PEEPfinder”.
The PEEPfinder can be started directly or, alternatively, via the optional pre-oxy
-
genation function with an adjustable oxygen concentration.
•
Upon starting the PEEPfinder, the system first determines the respiratory situ
-
ation (leakage, trapping, ...) and initiates a prolonged expiratory phase if needed.
•
Then, the quasi-static manoeuvre begins at the set starting point and proceeds
as configured (inspiration only, expiration only, or inspiration and expiration).
•
For the duration of the recruitment, between inspiration and expiration, the
valves are closed. At the end of the recruitment time, the expiratory valve
opens and the expiratory phase begins (no bias flow (Byflow) is added).
•
Upon termination of the PEEPfinder manoeuvre, ventilation continues.
Remark
If, during the manoeuvre, a situation occurs which makes it impos-
sible to determine a valid pressure situation (onset of spontaneous
breathing, high leakage, reaching the limit volume, high recruitment
effects, ...), the manoeuvre will be aborted and the graphic and
numeric evaluation will not be displayed.
Remark
The manoeuvre can be terminated prematurely at any time with
the “Stop” and “Enter” buttons.
Remark
Duration of the manoeuvre: 2.5 minutes maximum
Summary of Contents for elisa 300
Page 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Page 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Page 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Page 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Page 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Page 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Page 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Page 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...