8 EMC information
EN
77
Emission limits
Conducted and radia-
ted RF emissions
CISPR 11
Harmonic distortion
IEC 61000-3-2
Voltage fluctu-
ations/flicker
IEC 61000-3-3
Housing
Phenomenon
Test method
Immunity test levels
Electrostatic discharge IEC 61000-4-2
±8 kV contact
±15 kV air
Radiated RF
disturbances
IEC 61000-4-3
28 V/m
80 MHz to 6 GHz
80 % AM at 1 kHz
Magnetic field at
rated power
frequency (50 Hz)
IEC 61000-4-8
30 A/m
Power supply AC
Phenomenon
Test method
Immunity test levels
Fast transient electrical
disturbances/bursts
IEC 61000-4-4
±2 kV
100 kHz repetition
frequency
Impulse voltages/
surges
IEC 61000-4-5
±1 kV line-to-line
±2 kV line-to-ground
Conducted RF distur-
bances
IEC 61000-4-6
3 V
0.15 MHz to 80 MHz
6 V in ISM bands and amateur
radio bands between
0.15 MHz and 80 MHz
80 % AM at 1 kHz
Voltage dips
IEC 61000-4-11
0 % U
T
for ½ period
At 0°, 45°, 90°, 135°, 180°,
225°, 270°, 315°
0 % U
T
for 1 period
70 % U
T
for 25 periods
one-phase at 0°
Voltage
interruptions
IEC 61000-4-11
0 % U
T
for 250 cycles
Note: U
T
is the AC mains voltage prior to the application of the test levels.