MAT-1 Battery Replacement Tool Kit, Rev A
3
13.
Verify that the “Logger Time” has reset to the year 2000, that the battery status is “Good” and
that the “Real-time Data” is updating. If any of them are not true, then the logger was not fully
reset and the logger must be power-cycled again. Go back to step 2 and increase the time
between disconnecting and reconnecting the battery.
14.
You are done! Reset the logger’s clock and continue using your MAT-1 Data Logger.
Support
Additional information is available on the Lowell Instruments web site. Please see:
http://www.lowellinstruments.com/Support
for the
most up-to-date support information.
Warning
: This logger contains a lithium battery. Do not cut open, incinerate, heat above 85°C (185°F), or
recharge the lithium battery. The battery may explode if the logger is exposed to extreme heat or
conditions that could damage or destroy the battery case. Do not dispose of loggers or batteries in fire.
Do not expose the contents of the batteries to water. Dispose of the batteries according to local
regulations for lithium batteries.
Contact Information
Lowell Instruments, LLC
33 Cameron Rd.
North Falmouth, MA 02556
U.S.A.
Phone: (0011) 1-508-444-2616
Email:
© 2014 Lowell Instruments, LLC