10
TurbiDirect_6 02/2019
2.1 Operation
2.1.1 Commissioning
Before working with the TB300 IR insert the rechargeable batteries and the Lithium battery
(part of delivery). The rechargeable batteries are not charged. See chapter 2.1.2 Saving data
– Important Notes, 2.1.3 Replacement of rechargeable batteries resp. Lithium battery. and
2.1.4 Charging the rechargeable batteries.
Before working perform the following settings in the Mode-Menu:
• MODE 10:
Select language
• MODE 34:
perform „Delete data“
• MODE 12:
Set date and time
See chapter 2.4 Photometer settings.
2.1.2 Saving data – Important Notes
The Lithium battery saves data (stored results and photometer setting) if there is no power
from the power supply from the rechargeable batteries or the mains adapter.
Recommendation:
Exchange of the lithium battery every 5 years.
Note:
When neither mains adapter nor batteries supply energy to the instrument, all stored
data and settings will be lost, if the lithium battery is taken out.
Recommendation:
Keep the instrument connected to mains adapter supply while
changing the lithium battery.
2.1.3 Replacement of rechargeable batteries resp.
Lithium-battery
1. Switch the instrument off.
2. If necessary remove vial from the sample chamber.
3. Place the instrument upside down on a clean and even surface.
4. Unscrew the two screws (A) of the battery compartment cover (B).
5. Lift battery compartment cover off.
6. If necessary remove old rechargeable batteries (C) and/or the Lithium-battery (D).
7. Place 7 new rechargeable batteries and/or the Lithium-battery.
Ensuring the correct polarity!
8. Replace the battery compartment cover.
9. Tighten the screws carefully.
CAUTION:
Dispose the used batteries in accordance with all federal, state and local regulations.
2.1.4 Charging the rechargeable batteries
The rechargeable batteries are uncharged in the instrument. As soon the TB300 IR is
connected by the adapter to the mains the batteries are charged.
Empty rechargeable batteries should be charged in the instrument for at least 5 days.
10 charging and discharging cycles are necessary before the rechargeable batteries obtain
their full capacity.
It is possible to operate the instrument with the adapter with or without inserted
rechargeable batteries.