WARNING:   

Do not attempt to calibrate this device

with the cover removed.  Potentially lethal voltage is present

on some of the internal components.  Make sure that the

cover is firmly in place before plugging into the socket.

CALIBRATION

1.

Apply the appropriate input for the low end of the scale.

2.

Adjust the ZERO screw for the

appropriate low end output.

3.

Apply the appropriate input for

the high end of the scale.

4.

Adjust the SPAN screw for the

appropriate high end output.

5.

Repeat as necessary.

MICHIGAN CITY, IN

© COPYRIGHT 1995 DWYER INSTRUMENTS, INC., ALL RIGHTS
RESERVED                                                     PRINTED IN U.S.A. 3/95

949-1203

Dwyer Instruments, Incorporated

PO Box 338       Michigan City, IN 46361-0338

(800) 828-4588       (219) 879-8000       FAX (219) 872-9057

www.love-controls.com

LOVE  CONTROLS  DIVISION

®

SPECIFICATIONS

Power Supply:

85 to 265 VDC / VAC 50 to 400 Hz.

Isolation:

1500 VAC

Ambient Temperature Range (operating):

 0° to 55° C (32° to 131° F)

Linearity:

0.25%

Drift:

±0.02% per °C typical, ±0.05% maximum.

Maximum current output load:

600 ohms.

Maximum voltage output load:

20mA (500 ohms).

Input Characteristics:

Thermocouple Impedance:  3 Megohms minimum.
RTD search current:  500µA maximum.