Instruction / Description of the Operating Mode
5.1 Connect Description
Terminal assignment
1 Signal
0
±10V
Input
2 GND
Input
3 NC
-
4 10
...
30VDC
Supply
5 GND
Supply
6 Shield
Supply
7 Clear
min.
value
Input
8 GND
Input
9 Clear
max.
value
Input
10
Min.
value
Output
11
GND
Output
12
Max.
value
Output
Status display (LED)
Power:
On
during
operation
Delete
MAX:
On
during
deletion
Delete
MIN:
On
during
deletion
5.2 Switching the Device on
The device is activated and ready to operate as soon as supply voltage is applied. The readiness for
operation is signaled by the Power-LED.
5.3 Minimum-/Maximum Value Memory
A signal is measured at terminals 1 and 2. If the following measurement is smaller, the signal output at
the minimum value memory is updated at terminals 10 and 11. If the measurement is higher, the
respective value of the previous measurement is held and issued. The minimum value output is from
-10V to +10V.
A signal is measured at terminals 1 and 2. If the following measurement is larger, the signal output at
the maximum value memory is updated at terminals 12 and 11. If the measurement is smaller, the
respective value of the previous measurement is held and issued. The maximum value output is from
-10V to +10V.
A control signal of 10 ... 30VDC is required in order to reset the minimum and maximum values. In
order to delete the minimum value, the control signal must be applied to terminals 7 and 8. In order to
delete the maximum value, the control signal must be applied to terminals 9 and 8.
Power
GM 42
MAX
MIN
1 2 3
4 5 6
12 11 10
9 8 7
Bottom clamps
Top clamps
Bottom clamps
Top clamps
delete
delete