Figure 52 - Plot Trendline
6. The formula of the line is
y=mx+b , where y is the engineering units at a given point
(measurement),
m is the slope of the line that represents the linear ratio, x is the A/D value at
a given point, and
b is the fixed zero load offset of the sensor. Enter the slope and offset as
the conversion values for the sensor channel under the applicable engineering units. In this
example, enter 0.1338 for the slope and -282.36 for the offset for the units conversion values
on the measured channel.
Figure 53 - Slope and Offset Values
70