13 MAINTENANCE SOFTWARE
13.1 Calibrating the table sections to the zero position
Calibrating the OT340, you need computer for use the Maintenance
software. For connecting the OT340 and the computer, you need D9-D9
data cable.
Make sure that the table Stand by switch is Off position
Connect the computer to the service plug using a data cable
Push the Stand by switch On position
Open the Service software
Select the Calibration interleaf