background image

Operating instructions 

4040, 4050 and 4140 examination tables/Version 1.15 

Copyright © Lojer Oy, 2007 

24

 

 

11.1

 

International warranty conditions 

1. The warranty is only valid if, when warranty service is required, the warranty demand is 
presented with the original invoice or sales slip or (order) confirmation, and the serial 
number on the product has not been defaced. 
2.  Lojer’s  obligations  are  limited  to  the  repair or,  at  its  discretion,  replacement  of  the 
product or the defective part. For this, the customer is responsible for the despatch and 
insurance  of  the  product.  The  transport  and  insurance  costs  shall  be  borne  by  the 
customer.  Lojer  shall  assume  the  costs  for  the  materials  /  spare  parts  and  the  labour 
costs as required, as well as the return postage to the sender.  
3. In case of warranty repairs, they must be carried out by Authorised Lojer Dealers. No 
re-imbursement will be made for  repairs  carried out by  non Lojer representative  and, 
any such repair work and damage to the products caused by such repair work will not 
be covered by this warranty. 
4.  This  product  is  not  considered  to  be  defective  in  materials  nor  workmanship  by 
reason that it requires adaptation in order to conform to national or local technical or 
safety standards in force in any Country other than the one for which the product was 
originally designed and manufactured. 
5. This warranty covers none of the following: 
a) Periodic check-ups, maintenance and repair or replacement of parts due to normal 
wear and tear (such as upholsteries, ropes or reduced capacity of accumulators) 
b) Cost relating to transport, removal or installation of the product. 
c) Misuse, including the failure to use this product for its normal purposes or incorrect 
installation. 
d)  Damage caused by  Lightning,  Water,  Fire,  Acts  of  God, Natural  Catastrophes, War, 
Public Disturbances, incorrect mains voltage or any other cause beyond the control of 
Lojer. 
e)  Spillage  of  chemicals  or  liquid  or  use  of  any  other  hazardous  or  non-instructed 
substances, which may affect the product. 
f)  Minor  defects  or  deviations  from  the  product  specifications,  that  are  immaterial  or 
negligible or trivial as regards to the value or the functioning of the product 

11.2

 

Proceeding in a warranty case  

The  following  information  is  an  explanation  of  what  you  must  do  in  the  case  of  a 
warranty claim. At first, please contact your Lojer dealer. There our employees will help 
to  diagnose  defects  and  malfunctions.  They  are  often  capable  of  eliminating  the 
malfunction over the phone.  
Please always have the following information  
ready before calling the dealer:  

Summary of Contents for Afia 4140

Page 1: ...Afia 4040 4050 and 4140 Examination tables Operating instructions...

Page 2: ......

Page 3: ...ements of table top sections 7 5 2 Locking the wheels 8 5 3 Height adjustment 8 5 3 1 Height adjustment with hand control 8 5 3 2 Height adjustment with foot control 9 5 4 Adjusting the backrest 10 5...

Page 4: ...cushion 16 5 16 Mounting of infusion pole 16 5 17 Mounting of anaesthesia screen 18 6 Care and maintenance 18 6 1 Cleaning and care of Upholstery 18 6 2 Disinfection 19 7 Service 19 7 1 Half yearly m...

Page 5: ...n hospitals and clinics The models 4040 and 4140 have a two piece table top the 4050 has a three piece table top The models 4140 and 4050 have a gynaecological seat at the rear end of the table Read t...

Page 6: ...table 2 2 Afia 4050 1 Hand control 2 Paper roll holder 3 Mounting adapter 4 Infusion pole accessory 5 Arm support plate accessory 6 Anaesthesia screen accessory 7 Backrest 8 Gynaecolocical backrest 9...

Page 7: ...er Oy 2007 3 15 Bowl 16 Foot control 17 Central locking pedal Figure 2 Afia 4050 gynaecological general examination table 2 3 Afia 4140 1 Hand control 2 Paper roll holder 3 Backrest 4 Heel support 5 B...

Page 8: ...n 3 Table top up 4 Table top down Figure 4 Hand control functions 2 5 Options and accessories for examination tables Examination tables are available with following options Separately lockable wheels...

Page 9: ...Afia examination tables and couches The symbols are located on the control unit under the table top Symbol Description This product complies with guideline 93 42 EEC IPX4 Protection against fluids ele...

Page 10: ...oose between the plug and the examination table Do not tie up the cord to the examination table because the raising movement can cut it off NOTE Do not place anything for example a chair under the exa...

Page 11: ...ing nuts 3 pcs onto the screws from inside the table frame with the spanner 6 Tighten the screws with the Allen key Figure 5 Side rail 5 Using the examination table 5 1 Movements of table top sections...

Page 12: ...by pressing down the green end of the pedal by the wheel to be locked Make sure that the wheel is locked by moving the examination table sideways 1 Wheel immobilized red end down 2 Wheel free pedal in...

Page 13: ...press it or stand on it The table top rises by moving the foot control on rear end of the table upwards PUSH and goes down by moving the foot control down PULL Figure 9 Foot control on rear end of the...

Page 14: ...etween 0 and 78 1 Raise 2 Lower Figure 11 Adjusting the backrest with hand control 5 5 Adjusting the gynaecological backrest 4050 Gynaecological backrest middle section of examination table 4050 can b...

Page 15: ...an be lifted to two different angles 10 eller 15 1 Handle Figure 13 Adjusting the seat model 4050 5 7 Adjusting the heel supports 4140 and 4050 Examination tables 4140 and 4050 have longitudinally adj...

Page 16: ...e the bowl by swivelling it first out and lifting it from the holder Figure 15 Bowl 5 9 Paper roll holder The paper roll holder is located under the front end of the examination table The paper roll h...

Page 17: ...t the table top The inclination is locked when the handle is released 1 Trendelenburg handle Figure 17 Trendelenburg function 5 11 Mounting of mounting adapter The mounting adapter is mounted onto the...

Page 18: ...e 5 12 1 Mounting of arm support 1 Push the arm support stem pin through the mounting adapter hole 2 Tighten the arm support into its position with the mounting adapter screw 5 12 2 Removing of arm su...

Page 19: ...middle of the rail 1 Hold the hand rail from its upper tube with one hand 2 Lift the locking knob 3 Lower the hand rail simultaneously The rail can be raised by lifting it up The rail locks in positi...

Page 20: ...ightening height and width by pulling out and turning the handle 5 15 2 Inclination adjustment of the leg support cushion 1 Loosen the adjuster wheel under the leg support cushion 2 Tip up the cushion...

Page 21: ...Operating instructions 4040 4050 and 4140 examination tables Version 1 15 Copyright Lojer Oy 2007 17 1 Infusion pole 2 Mounting adapter Figure 22 Infusion pole...

Page 22: ...ssible with lukewarm water and a damp cloth The use of a common Microfiber cloth is especially recommended for this purpose In case of heavy soiling we recommend using a mild cleaning agent possibly 3...

Page 23: ...operations on the examination table NOTE Always remove the power cord from the plug before servicing the examination table 7 1 Half yearly maintenance Check the condition of the power cord the spindle...

Page 24: ...s follows 11 Remove the filler plug If there are no filler plug and hole drill a hole with 6 mm drill in pre marked recess hole The adjusting screw is inside 12 Adjust the brakes with a screwdriver 13...

Page 25: ...edical devices its revisions and enforcing national regulations The following conformable standards have been applied in designing and testing the products SFS EN 292 2 SFS EN 349 SFS EN 60601 1 IEC 6...

Page 26: ...d replace if necessary For changing a motor hand or foot control or control unit contact Lojer service see Section 12 Contact information Please always have the following information ready before call...

Page 27: ...quality product from Lojer Our products are developed to meet high standards in both quality and technology To further ensure quality we grant you the final customer warranty on our products in accord...

Page 28: ...orm to national or local technical or safety standards in force in any Country other than the one for which the product was originally designed and manufactured 5 This warranty covers none of the foll...

Page 29: ...ne number at which you can be reached The model designation the article code and the serial number of the product The date of purchase A brief description of the problem 12 Contact information Manufac...

Reviews: