Operating instructions
4040, 4050 and 4140 examination tables/Version 1.15
Copyright © Lojer Oy, 2007
19
6.2
Disinfection
You can disinfect the parts of the examination table for example with ERISAN Des
disinfection agent (available from Lojer sales, see Section 12
Contact information.
See
the package for dilution instructions)
NOTE!
ERISAN Des disinfection agent is not suitable for synthetic leather
upholsteries.
NOTE!
Disinfection agents wear out the cushion surface. Always wipe away the
disinfection agent with a damp cloth.
7
Service
NOTE! Only professional personal may carry out service and repair operations
on the examination table.
NOTE! Always remove the power cord from the plug before servicing the
examination table.
7.1
Half-yearly maintenance
Check the condition of the power cord, the spindle motors and hand control cables and
replace if necessary.
7.2
Yearly maintenance
Check and lubricate the following parts once a year or more often if necessary.
Joints
Bearings
Bearing points of the underside rods.
Lubricate with e.g. Würth HHS 2000 synthetic vaseline spray.
7.3
Adjusting the brakes
Basic adjustment of the central locking plastic wheels is done as follows. Individually
locking wheels cannot be adjusted.
4.
Lift the examination table on a stand or on a chair. The wheel to be adjusted must be
aloft.
5.
Remove the fixing screws (2) of the wheel.
6.
Turn the wheel body half a turn at a time.