Operating the Logger
UTRED-16 User Guide
23
Here, the temperature was below the secondary alarm threshold, but above the tertiary
threshold; an alarm event has not been triggered:
An alarm was triggered
If an alarm event occurred the alarm indicator is
displayed in the upper left corner of the display.
Additionally, one of the alarm limit markers or is
displayed, to show if the event was an upper or a lower
alarm.
The alarm marker
in this screen does not indicate which alarm threshold
caused the alarm, only that the most recent temperature recorded, which you see
on the display (18.8 °C) was above the secondary threshold.
To identify which of the alarm conditions caused the alarm event you will need to
review the data on screen as described in
Reviewing Alarms on the next page
, or
open the PDF file on your PC.
Marking a reading with an inspection mark
When you press the button while the UTRED-16 is recording, the next reading taken will
be identified in the downloaded data and on the PDF report with an inspection mark.
The
MARK
symbol appears and remains shown until the next reading is taken.
If the
Paused Readings
function is enabled, you will also see the
PAUSED
symbol.
An inspection mark will also be recorded if you clear an alarm.