background image

LogTag

® 

 - 

20 Day Pharmaceutical Temperature Indicator with display

 

© Copyright 2013, LogTag Recorders Limited 

page 16 of 17 

TIC20 Factory Pre-sets 

LogTag

®

 TIC20 indicators cannot be client configured, but instead are purchased with one of a 

number of different factory configuration profiles pre-installed. If a default configuration does not suit, 
customers can compile their own profile by specifying the parameters in the table below when 
ordering. Please note that minimum order quantities apply for TIC20 units with profiles not stocked by 
LogTag

®

The default configuration profile is called “Profile 0006”; each custom configuration is assigned a 
different profile number which can be used for ordering. 

 

Description 

Profile 0006

 

(default)

 

Range / Option 

Requirement 

Maximum Days 

(Maximum number of days to monitor)

 

20 

1-20 

 

Sampling interval 

6 minutes* 

1 to 15 minutes 

 

Start Delay 

60 minutes 

0 to 99 minutes, 0=no delay 

 

Display Temperature units 

°C 

°C or °F ** 

 

Hibernate once STOPPED 

(Stops the Real Time Clock, increases length of period in 
which the unit can be downloaded)

 

off 

on or off 

 

Allow stopping with STOP button 

on 

on or off 

 

ALM 1 

(Alarm 
Limit 1)

 

Monitor this alarm 

enabled 

enabled or disabled 

 

Temperature limit value 

10°C 

-30°C to 60°C (-22°F to 140°F) 

 

Alarm activation 

instant 

instant, accumulated or consecutive*** 

 

Activation delay time**** 

N/A 

1 minute to 45 days 

 

Direction 

upper 

upper/lower 

 

ALM 2 

(Alarm 
Limit 2)

 

Monitor this alarm 

enabled 

enabled or disabled 

 

Temperature limit value 

8°C 

-30°C to 60°C (-22°F to 140°F) 

 

Alarm activation 

accumulated 

instant, accumulated or consecutive*** 

 

Activation delay time**** 

12 hours 

1 minute to 45 days 

 

Direction 

upper 

upper/lower 

 

ALM 3 

(Alarm 
Limit 3)

 

Monitor this alarm 

enabled 

enabled or disabled 

 

Temperature limit value 

2°C 

-30°C to 60°C (-22°F to 140°F) 

 

Alarm activation 

accumulated 

instant, accumulated or consecutive*** 

 

Activation delay time**** 

12 hours 

1 minute to 45 days 

 

Direction 

lower 

upper/lower 

 

ALM 4 

(Alarm 
Limit 4)

 

Monitor this alarm 

enabled 

enabled or disabled 

 

Temperature limit value 

0°C 

-30°C to 60°C (-22°F to 140°F) 

 

Alarm activation 

instant 

instant, accumulated or consecutive*** 

 

Activation delay time**** 

N/A 

1 minute to 45 days 

 

Direction 

lower 

upper/lower 

 

 

6 minutes = 0.1 hour; preferred intervals are 1, 2, 3, 6 and 12 minutes 

** 

When °F is chosen, temperatures above 99°F will be shown on the display without a decimal point, but will be listed correctly in downloaded 
data. 

***  Instant = one temperature reading is above/below limit; consecutive = temperature readings are above/below limit for the time defined in the 

activation delay time without interruption (single event); accumulative =  temperature readings above/below limit for a total of time defined in 
the Activation delay time, but may not necessarily be sequential. 

****  Delay times for alarm activations must be in multiples of the sampling interval. 

Summary of Contents for TIC20

Page 1: ...LogTag 20 Day Transit Temperature Indicator with display DOCUMENT REVISION 1 3 16 Apr 2013 Copyright 2013 LogTag Recorders Limited...

Page 2: ...MJ Mod to TIC20 20 day device Updated LCD images and added the four WHO types 0 3 7 12 2011 MJ Updated LCD images to include triggered ALARM temperature points during MONITORING and STOPPED modes 0 4...

Page 3: ...downloaded if required via the free LogTag Analyzer companion software and the TICT TIC20 interface Two push buttons allow starting and stopping the Indicator and can also be used to navigate the vari...

Page 4: ...monitoring Current Temperature Monitored Load Load OK For a reject load will appear with additional information about the Alarms Battery Condition Battery OK For Low battery will appear Recording Summ...

Page 5: ...when the indictor has been stopped can the statistics be displayed Limit Markers For each day statistics for up to four different temperature limits are collected If an arrow is displayed next to ALM...

Page 6: ...K Low After 30 seconds the Indicator will re enter a hibernated state and the display will switch off To start the Indicator press and hold the START button for 5 seconds The MONITORING indicator will...

Page 7: ...ring Display LOAD REJECT In this example the device has been monitoring for 6 days and 1 hour and is still active The symbol has been replaced by the symbol and since a 45 C ALARM has occurred the loa...

Page 8: ...pending stop Stopping Display pending stop waiting for release When STOPPED is permanently lit release the STOP button within 2 seconds If you release the STOP button before the STOPPED text becomes p...

Page 9: ...ed page 9 of 17 Battery indication The battery condition can be OK or LOW as shown in the display examples below An Indicator should not be used if it shows the low battery symbol at the beginning of...

Page 10: ...tinuously if any alarm was triggered during monitoring Temperature statistics for day 5 are displayed cursor next to The temperature went below the 0 5 C limit next to ALM1 which also triggered the AL...

Page 11: ...no n a 0 30 deg 2 0 2 no n a 2 0 5deg 0 0 0 no n a 18 DAY 3 45 deg 0 0 0 no n a 0 30 deg 2 0 2 no n a 2 0 5deg 0 0 0 no n a 0 DAY 4 45 deg 0 0 0 no n a 0 30 deg 2 0 2 no n a 2 0 5deg 0 0 5 5 no n a 0...

Page 12: ...arm as the alarm trigger requires cumulative exposure of 10 hours or more to temperatures above 30 C Press and release the REVIEW button to advance to the next screen On day 1 temperatures above 45 C...

Page 13: ...creen now shows data for day 4 indicated by the arrow next to Temperatures below 0 5 C were recorded for a time of 30 Minutes indicated by the next to ALM1 This did not trigger an alarm as only consta...

Page 14: ...ing displayed and no alarms have been triggered 7 Press and release the REVIEW button Since no further events were recorded for day 6 the screen now shows data for day 7 indicated by the arrow next to...

Page 15: ...in at hour 12 but only the 1 st alarm statistic is shown 9 Press and release the REVIEW button Since no further events were recorded for day 8 the screen now shows data for day 9 indicated by the arro...

Page 16: ...e Activation delay time N A 1 minute to 45 days Direction upper upper lower ALM 2 Alarm Limit 2 Monitor this alarm enabled enabled or disabled Temperature limit value 8 C 30 C to 60 C 22 F to 140 F Al...

Page 17: ...7 minutes by method detailed in EN12830 1999 Vibration Withstands vibration specification as detailed in EN12830 1999 Shock Withstands shock specification as detailed in EN12830 1999 Withstand 5 drops...