background image

Data from 7 days ago

After pressing the 

 button for a few times (skipping days -3 to -6) 

the maximum statistic from seven days ago is displayed:

 

l

The 

 flashes and 

-07

 

DAYS

 is shown to indicate that the data 

being displayed is from seven days ago.

 

l

Seven days ago recordings above the upper threshold were recorded, 
indicated by the upper alarm threshold marker

 

l

The duration was 6 hour 30 minutes, which was shorter than the 
allowed period, so no alarm event was generated 

 

l

The maximum temperature recorded seven days ago was 10.5°C.

Data from 8 days ago

After pressing the 

 button twice (skipping the minimum statistics 

for day -7) the maximum statistic from eight days ago is displayed:

 

l

The 

 flashes and 

-08

 

DAYS

 is shown to indicate that the data 

being displayed is from eight days ago.

 

l

Eight days ago recordings above the upper threshold were recorded, 
indicated by the upper alarm threshold marker

 

l

The duration was 5 hour 30 minutes

 

l

An alarm event was generated  .  Neither duration from day -7 or -8 
by itself would cause an alarm event, however the combined 
duration is 10 hours 50 minutes, which is more than the allowed 10 
accumulative hours.

 

l

The maximum temperature recorded eight days ago was 11.5°C.

Tip: Pressing and holding the 

 button will get you back to the 

previous screen, rather than advancing one screen.

Stopping the TRED30-16R

Automatically

The TRED30-16R automatically stops recording temperature when the 
maximum number of readings specified during configuration has been 
reached. Your unit can also be set up to stop automatically when it is 
downloaded. This option needs to be set up at the factory and cannot 
be changed during configuration with LogTag

®

 Analyzer. Your 

distributor can supply more information about this option.

Manually 

You can configure a TRED30-16R so it can be stopped with the 

START

/CLEAR/STOP

 button. This feature is useful when you take the 

recorder out of a shipment and don’t want to falsify the statistics with 
readings taken after the shipment completion. The stop function  is 
enabled in the 

Advanced Options

 dialogue during configuration (see 

Advanced Configuration Options 

 on page 6). When enabled, following 

will stop the unit (quickstart enabled):

Press and hold the 

START

/CLEAR/STOP

 button.  

STOPPED

 is 

shown in addition to the  

 symbol.

Then  the 

 symbol disappears:

Once the 

 symbol disappears, release the button within 2 seconds. 

The recorder will now stop taking readings.

If the quickstart option is disabled, 

STOPPED

 will flash for approx. 6 

seconds. When the symbol remains permanently on, release the button 
within 2 seconds. The  

 symbol now disappears and the recorder is 

stopped.
 
The recorder will 

not 

stop if you…

Page 14 of 17

TRED30-16R Product User Guide, Revision 1.4, copyright © 2004-2017,

LogTag

®

Recorders  

. All rights reserved.

Distributed by MicroDAQ            www.MicroDAQ.com           (603) 746-5524

Summary of Contents for TRED30-16R

Page 1: ...corder with Display and 30 Day Statistics Memory Product User Guide Document Release Version 1 4 Published 1 March 2017 Copyright LogTag Recorders 2004 2017 LogTag Recorders Distributed by MicroDAQ www MicroDAQ com 603 746 5524 ...

Page 2: ...aused Readings 12 Sensor is disconnected 12 Power Save 12 Total Number of Days Statistics Collected 12 Reviewing Day Statistics on the TRED30 16R 13 Stopping the TRED30 16R 14 Automatically 14 Manually 14 Resetting the Recorder 15 Hibernating a TRED30 16R 15 Technical Specifications 16 Appendix 1 Battery Replacement 17 Revision History Version Date Details 1 0 9 12 09 Initial release 1 1 3 5 10 Ch...

Page 3: ...readings if the device was used with a low battery l For consequential loss Useful Life The operational life of the TRED30 16R is approx 12 months after configuration provided the device was not stored for over 12 months prior to activation the device is not downloaded excessively more than once a week data is reviewed on the display no more than once daily for 30 seconds The recording interval is...

Page 4: ...the statistical data directly on the display It is also used to place an inspection mark in the data listing Display The display shows at a glance if alarm events have occurred for both the current day and up to 29 days in the past Details of any alarm event can be checked directly on the unit by inspecting the statistics history on the recorder s display or in more detail by downloading the logge...

Page 5: ... be instant consecutive or accumulative1 and a delay time if it is not an instant alarm If an alarm trigger condition requires readings to exceed an upper threshold temperature it is called an upper alarm If an alarm trigger condition requires readings to go below a lower threshold it is called a lower alarm The alarm triggering structure in the recorder supports many different alarm trigger confi...

Page 6: ...anced Configuration Options Select Advanced Options for additional configuration settings These settings decide how some of the elements are displayed on the screen and set certain options specific to the TRED30 16R Figure 4 TRED30 16R advanced configuration screen Parameters that influence the appearance of the display are l Pause alarm statistics processing see Paused Readings on page 12 l Tempe...

Page 7: ...er l If is shown together with the word PAUSED the product is also recording but the recorded values are not taken into account when calculating alarm events and durations l If the word STOPPED is shown the TRED30 16R has finished recording temperature data Time Value and Time Indicators The time value display is used to show one of the following l a clock with the current time l time remaining to...

Page 8: ... the clock press and hold the button then press and hold the button Keep holding both buttons together continuously for a period of 8 seconds Press the buttons firmly During this period the CLOCKADJ icon flashes Release the buttons when the flashing stops The clock can now be adjusted Initially the minutes digits flash Press the button to increment the minutes digit once it reaches 59 it rolls bac...

Page 9: ...2 seconds after READY disappears Push button start quickstart disabled Again the recorder must be in READY mode for it to be started Press and hold the START CLEAR STOP button The STARTING symbol starts flashing for about 4 seconds Then the symbol remains on permanently for 2 seconds If you release the button during this time The symbol is shown the STARTING symbol goes off and the recorder starts...

Page 10: ...ear and READY will show together with the current time and the recorder can now be started again as normal The TRED30 16R will not be re set if you release the button before STARTING disappears or if you keep holding the button until STOPPED disappears Automatic date time start If you configured the TRED30 16R for a date time start it will start recording temperature values as soon as the entered ...

Page 11: ...xt reading taken by the TRED30 16R will be identified in the downloaded data with an inspection mark An inspection mark will also be recorded if you clear an alarm Clearing an Alarm During configuration with LogTag Analyzer you can allow a user to clear an alarm on the display This is a useful function for an inspector so new alarms can be recognised easier This display screen shows an existing al...

Page 12: ...onnected for 3 hours during this time no alarm would be generated After downloading the TRED30 16R with LogTag Analyzer any readings taken while the sensor is disconnected are shown as in the data list The graph will show a gap during this period If the sensor was disconnected for a complete day the minimum and maximum statistics for that day will show on the display and on the list Power Save Whe...

Page 13: ... above the limit yesterday which was 12 hours 30 minutes l an alarm was triggered indicated by the as this duration was longer than the allowed time above the threshold l The maximum temperature recorded yesterday was 12 9 C Pressing the button now displays yesterday s minimum statistic l still flashes and 01 DAYS still shows as we are still looking at yesterday s data l Yesterday no temperature v...

Page 14: ...creen Stopping the TRED30 16R Automatically The TRED30 16R automatically stops recording temperature when the maximum number of readings specified during configuration has been reached Your unit can also be set up to stop automatically when it is downloaded This option needs to be set up at the factory and cannot be changed during configuration with LogTag Analyzer Your distributor can supply more...

Page 15: ...R via LogTag Analyzer see Advanced Configuration Options on page 6 Hibernating a TRED30 16R When hibernated the recorder s power consumption is near zero and the life of the battery approaches its shelf life typ 5 or more years This is useful for conserving battery life when the recorder is not used for extended periods TRED30 16R recorders are placed into Hibernation using LogTag Analyzer by clic...

Page 16: ...e start up to 180 days in the future 1 Values above 199 9 F cannot be shown on the display The values will however be recorded and shown in the LogTag Analyzer software Alarm functions l one configurable upper alarm l one configurable lower alarm l OK tick and Alarm cross on display linked to alarms l Audible alarm Vibration Withstands vibration specification as detailed in EN12830 1999 Shock l Wi...

Page 17: ... temperature range of the battery and make sure it covers the range of the TRED30 16R l Put a new CR2032 cell into the battery compartment as shown the right hand edge of the battery is inserted first against the contact l Press firmly down on the left hand side of the battery to click the battery into place l Replace the battery door turn clockwise to lock it into place l Cover the battery door w...

Reviews: