4
2. Buffer reservoir temperature control
a. Fill the buffer reservoir with 1.2 L of the Electrophoretic Tissue Clearing Solution.
b. Screw on the cap of the buffer reservoir making sure the temperature probe and buffer reservoir
outlet tube are both in the reservoir.
c. Insert the temperature probe connector that projects from the cap of the X-
CLARITY™ buffer
reservoir into the temperature probe port on the rear panel of the X-
CLARITY™ ETC Controller.
Check the home on the connector and the thickness of the two pins for proper orientation.
Reservoir outlet
Reservoir outlet tube
Temperature probe connector
Temperature probe
Reservoir inlet
Thick pin
Thin pin
Homing groove