The m

agnetic

 sensor detects whether the safety cover is open or closed. The X-CLARITY

 

Tissue Clearing System will not run if the safety cover is open even if you push the RUN/STOP

button. No electric current will enter the chamber if the safety cover is not closed.

5.

DRAIN 

the X-CLARITY

™ ETC Chamber and tube attached to its outlet before you open it again.

Turn the power off via the X-

CLARITY™ ETC Controller.

Change the direction of the pump to pump counterclockwise.

The tube connecting the X-CLARITY

 ETC Chamber outlet and the buffer reservoir outlet will

begin to empty into the X-CLARITY

 ETC Chamber. 

****Make sure this tube is completely

empty to prevent buffer spilling when you disconnect the tube from the chamber outlet.****

The buffer will flow out of the chamber inlet back into the reservoir. You can watch the buffer

level drop through the window of the X-CLARITY

 ETC Chamber.

When the X-CLARITY

 ETC Chamber is sufficiently drained, turn off the pump. Disconnect the

snap-lock connector tube from the chamber outlet. Open the safety cover. Unscrew the

chamber cap. Check your sample.

M

agnetic

 sensor

Summary of Contents for C10101

Page 1: ...ber X CLARITY ETC Controller Quick Start Guide X CLARITY TISSUE CLEARING SYSTEM This Quick Start Guide provides instructions to install and use the X CLARITY TISSUE CLEARING SYSTEM For detailed instru...

Page 2: ...12103 Snap lock connector tube clear tube 1 C12104 Peristaltic pump tube light yellow tube 1 C12002 Container holder holds the tissue container 5 C12001 Tissue container holds a whole adult mouse brai...

Page 3: ...Check the orientation of the pins before inserting the ends of the multi cable into the X CLARITY ETC Controller or X CLARITY ETC Chamber There are homing grooves on the ends to ensure proper orientat...

Page 4: ...be are both in the reservoir c Insert the temperature probe connector that projects from the cap of the X CLARITY buffer reservoir into the temperature probe port on the rear panel of the X CLARITY ET...

Page 5: ...Chamber Make sure there is a tight seal to prevent buffer leakage b The buffer reservoir outlet protrudes from the buffer reservoir cap with tubing that extends into the buffer reservoir alongside th...

Page 6: ...ing the lever to the left d Lift the clamps on both sides of the pump head Load the middle of the tube into the pump head between the rollers and the compression block Reservoir outlet Chamber inlet P...

Page 7: ...ever to the right Make sure the tube is secure 4 Sample insertion a Open the red safety cover of the X CLARITY ETC Chamber b Unscrew the cap to the X CLARITY ETC Chamber Lever to the right Safety cove...

Page 8: ...to prevent leakage Make sure to keep the platform stage around the holder dry to prevent electric shortages and keep from damaging the chamber e Carefully shut the red safety cover of the X CLARITY E...

Page 9: ...E THE CHAMBER SAFETY COVER WITH THE TUBE IN PLACE b Attach the other end of the tube to the buffer reservoir inlet 6 Connecting the power cords a Plug in the power cord to the pump Connect to a power...

Page 10: ...10 Buffer Flow X CLARITY buffer reservoir buffer reservoir outlet chamber inlet X CLARITY ETC Chamber chamber outlet reservoir inlet X CLARITY buffer reservoir...

Page 11: ...r supports constant current mode only d For more details please refer to the user manual 3 Turn on the power for the pump 4 Set up the pump s settings a Set to 30 50 rpm to begin circulating the Elect...

Page 12: ...ITY pump to fill or drain the X CLARITY ETC Chamber See the user manual for more details 4 BEWARE of electric shock Keep the X CLARITY Tissue Clearing System as dry as possible If buffer spills or lea...

Page 13: ...mp counterclockwise The tube connecting the X CLARITY ETC Chamber outlet and the buffer reservoir outlet will begin to empty into the X CLARITY ETC Chamber Make sure this tube is completely empty to p...

Page 14: ...t Blank Logos Biosystems Inc STE 930 Doosan Venture Digm 415 Heungan daero Dongan Gu Anyang Si Gyeonggi Do 431 755 Republic of Korea Tel 82 31 478 4185 Fax 82 31 478 4184 Email sales logosbio com Webs...

Reviews: