The m
agnetic
sensor detects whether the safety cover is open or closed. The X-CLARITY
™
Tissue Clearing System will not run if the safety cover is open even if you push the RUN/STOP
button. No electric current will enter the chamber if the safety cover is not closed.
5.
DRAIN
the X-CLARITY
™ ETC Chamber and tube attached to its outlet before you open it again.
Turn the power off via the X-
CLARITY™ ETC Controller.
Change the direction of the pump to pump counterclockwise.
The tube connecting the X-CLARITY
™
ETC Chamber outlet and the buffer reservoir outlet will
begin to empty into the X-CLARITY
™
ETC Chamber.
****Make sure this tube is completely
empty to prevent buffer spilling when you disconnect the tube from the chamber outlet.****
The buffer will flow out of the chamber inlet back into the reservoir. You can watch the buffer
level drop through the window of the X-CLARITY
™
ETC Chamber.
When the X-CLARITY
™
ETC Chamber is sufficiently drained, turn off the pump. Disconnect the
snap-lock connector tube from the chamber outlet. Open the safety cover. Unscrew the
chamber cap. Check your sample.
M
agnetic
sensor