USE AND MAINTENANCE MANUAL
– DLC
NET
Series
Use as a batcher
LOGIC S.r.l. - M0145Db.docx
Mod. L0006A01 of 30/05/2016
Page 69 of 78
6.4 Dosing management
6.4.1
Introduction
The instruments of the DLC
NET
series allow the dosage up to 4 automatic and
manual components for a total of up to 14 steps of the recipe.
At each step, the operator selects to dose one of the 4 automatic components or a
manual component.
An automatic component can be dosed more times in the same recipe.
In the second case, the instrument manually increases the index of manual
components, so as to keep track of their number.
In the first dosage, automatically the system will start the recipe processing. If you
have a recipe in progress, the
led is blinking on.
The operator can then perform a series of dosages, automatic or manual, where
the sequence and dosing quantity will be stored for printing.
Once the entire sequence is complete, you can finish the recipe and print the
report of the whole process, including start time and end time recipe.
6.4.2
Dosing of automatic
components (
)
To dose the automatic components follow these steps:
Press the
button to enter the ingredients selection menu (
).
The display will show the written (
), corresponding to the first
automatic ingredient.
Use the
button to select the desired automatic ingredient and press
to confirm.
Press
at any time to cancel and return to the previous menu.
The display shows
with the first digit blinking and waits for the
setpoint of the dosed.
Use the
button to increase the digit and the
button to switch
to editing the next one and press
to confirm.
Now the display shows the flashing
written, awaiting
confirmation that actually start the dosage.
The instrument then checks the following conditions:
o
Scale zero (
, see section 6.2.1)
o
Maximum scale capacity (
, see section 6.2.2)
If either one of the two conditions is not verified,
or
appears on the screen, according to the fault detected. Make the