Load cells digital interface with indicator

 

 
 

USE AND MAINTENANCE MANUAL 

 

This manual consists of 78 pages, annexes excluded. 

 
 
 

 
 
 
 
 

INSTRUMENT DATA 

Manufacturer 

LOGIC S.r.l. 

Via del Tovo 5/3 

I-27020 Travacò Siccomario (PV) 

Italy

 

Series 

DLC

NET

 

Typology 

Load cells digital interface with indicator 

 

Summary of Contents for DLC NET Series

Page 1: ... n ORD 31473 n SOC 4719 n FASC 6024 CCIAA 154531 Pagina 1 di 78 Mod L0006A01 del 30 05 2016 Doc n M0145Db docx DLCNET SERIES Load cells digital interface with indicator USE AND MAINTENANCE MANUAL This manual consists of 78 pages annexes excluded INSTRUMENT DATA Manufacturer LOGIC S r l Via del Tovo 5 3 I 27020 Travacò Siccomario PV Italy Series DLC NET Typology Load cells digital interface with in...

Page 2: ...USE AND MAINTENANCE MANUAL DLCNET Series INDEX LOGIC S r l M0145Db docx Mod L0006A01 of 30 05 2016 Page 2 of 78 INTENTIONALLY BLANK PAGE ...

Page 3: ... REVIEW Rewiev Date Description 1 0 10 10 2015 Initial release 1 1 08 04 2016 General bugs correction 1 2 02 05 2016 Printer connections described Beka RS232 Printer RS485 connection removed A 15 07 2016 New indexing review B 02 08 2016 Description of printing configuration in Batch mode and watchdog behavior description added Connection of start stop button added C 03 10 2016 Configuration passwo...

Page 4: ...USE AND MAINTENANCE MANUAL DLCNET Series INDEX LOGIC S r l M0145Db docx Mod L0006A01 of 30 05 2016 Page 4 of 78 BLANK PAGE ...

Page 5: ... Principle of operation 13 2 7 3 Technical details 14 2 7 4 Operating environment 14 2 7 5 Electromagnetic compatibility 14 3 Installation 15 3 1 Preliminary checks 15 3 2 Mechanical installation 15 3 3 Board connections 17 3 3 1 Power supply 17 3 3 2 Load cell 17 3 3 3 Ground connections 17 3 3 4 Data connection 18 3 4 Power supply connection 18 3 4 1 Layout of the terminal blocks 18 3 4 2 Power ...

Page 6: ...0 mA or 4 20 mA 32 3 9 RS 232 connection for printer 34 3 10 Parallel interface connection 35 3 10 1 Operating mode 35 3 10 2 Connection for parallel interface type PNP 36 3 10 3 Connection for parallel interface type NPN 36 4 Settings 39 4 1 Preliminary operations 39 4 1 1 Introduction 39 4 1 2 Setting speed ADC 39 4 1 3 Setting termination resistance 39 4 2 Fieldbus communication 40 4 2 1 PLC ha...

Page 7: ...DP address configuration 56 5 5 9 Real time clock configuration 56 5 6 Setpoint use and configuration 57 5 6 1 Setpoint intervention mode 57 5 6 2 Setpoint operating value 58 6 Use as a batcher 61 6 1 Working specification 61 6 1 1 General functionality 61 6 1 2 Dosing control 61 6 2 Batcher configuration 63 6 2 1 Scale zero configuration 63 6 2 2 Maximum capacity configuration 64 6 2 3 Auto corre...

Page 8: ...MAINTENANCE MANUAL DLCNET Series INDEX LOGIC S r l M0145Db docx Mod L0006A01 of 30 05 2016 Page 8 of 78 7 Decommissioning and dismantling 73 7 1 General indications 73 7 2 Procedure 73 8 FIGURES 75 9 TABLES 77 ...

Page 9: ...onservation of the manual The purpose of this manual is to provide the directions to use the instrument safely and perform the installation procedures and routine maintenance Any calibration adjustment and extraordinary maintenance operations are not covered in this text being the exclusive competence of the technical assistance which must assist on the instrument in accordance with the technical ...

Page 10: ...it is expected wear caused by the use of the instrument In no event LOGIC S r l shall bear the cost of charges for failure of production LOGIC S r l is liable for defects resulting from normal use of the machine and not those resulting from Incorrect installation Lack of maintenance or calibration Improper use of the instrument Manumission modification maintenance by the customer without the writt...

Page 11: ...t described above or shown in this manual 2 6 Instrument identification 2 6 1 Manufacturer data LOGIC S r l Via del Tovo 5 3 I 27020 Travacò Siccomario PV Italy 2 6 2 Marking The instrument shows the following screen printed marking on the case Figure 1 Equipment back serigraphy On the chassis the serial number is indicated Please communicate it to LOGIC S r l in case of malfunctioning or replacem...

Page 12: ...29 th 2014 and subsequent modifications L apparato elettronico in oggetto è conforme agli standard armonizzati di seguito elencati This electronic apparatus meets the requirements of the following harmonized standard EN 61000 4 2 Compatibilità elettromagnetica Prove di immunità a carica elettrostatica Electromagnetic Compatibity Electrostatic Discharge Immunity Test EN 61000 4 3 Compatibilità elet...

Page 13: ...mber of Profinet nodes is limited by application In addition to these functionality the board can be interfaced with a remote display as a weight repeater to be installed in safe or hazardous area through a serial port RS 485 in addition to the fieldbus connection 2 7 2 Principle of operation The DLC NET module is a load cells signal amplifier The supply voltage of the transducer is 5 Vdc with the...

Page 14: ...32 64 averaged values Protection IP20 Digital outputs EB board n 8 output 24Vdc 100mA total with all outputs on Serial outputs RS 485 and RS 232 on EB board Analog output EB board 0 20mA 4 20mA configurable Dimensions 30 x 135 x 107 mm W x H x D 2 7 4 Operating environment The environmental license plate requirements for the correct functioning of the instrument are Operating temperature 0 45 C St...

Page 15: ... with all accessories essential for mechanical and electrical installation 3 2 Mechanical installation The instrument in question is supplied in nickel plated aluminum case for DIN rail mounting Figure 3 shows the mechanical drawing of the instrument Figure 3 Mechanical drawing of DLC NET The instrument must be installed in a safe area in accordance with ATEX Directive The instrument includes a ca...

Page 16: ...M0145Db docx Mod L0006A01 of 30 05 2016 Page 16 of 78 Figure 4 DLC NET WM mechanical drawing DLC NET DIN has a 20 poles connector for an external keyboard DLC NET KB and can be installed vertically on a DIN rail as well Figure 5 Figure 5 DLC NET DIN mechanical drawing ...

Page 17: ...ompensation on the signal branches The junction boxes JBOX series have been designed for this purpose and properly solve the problem The maximum cable length between the junction box and the instrument depends on the specific situation of electrical noise in the plant which potentially can induce noise on the signal cable In normal situations cable lengths up to 300 meters can be used without prob...

Page 18: ... noise cables connected to the actuators motors etc 3 4 Power supply connection 3 4 1 Layout of the terminal blocks The instruments of DLC NET series have three terminal blocks connection with the Fieldbus connector which can be a D SUB 9 pin connector for the Profibus DP protocol or double RJ45 Ethernet port with integrated switch in the case of connection in Profinet Figure 6 Figure 7 Figure 6 L...

Page 19: ...present the led F which indicates a possible fault of the Fieldbus communication Concerning DIN rail mounting inside a control panel board terminals are set as follows Figure 8 DLC NET WM terminals model with on board keyboard Figure 9 DTR NET WM terminals model with external keyboard Where TB7 is a 20 poles connector reserved for keyboard connection ...

Page 20: ...commends the use of power supplies with standard current limitation eg 24V 1A with Icc 1 2 A With power supplies provided with intervention at short circuit type Fold Back or similar are possible problems The instrument in question in fact absorb a strong current at power for a limited time about 4 A for 10 ms Figure 10 Power supply connection If the instrument outputs are connected to the dosage ...

Page 21: ...ording to the following table Table 4 Load cells connection Terminal Color code CAB TS6 Connection TB1 1 GND Shld cable shield TB1 2 Red excitation TB1 3 Green sense TB1 4 Blue excitation TB1 5 Grey sense TB1 6 White signal TB1 7 Yellow signal Figure 11 Load cells connection 6 wires For connections with length less than 8 meters it is possible use the four wire connection shown below Figure 12 Loa...

Page 22: ...gh a barrier In case of use of a passive barrier below is an example of connection with the Zener barrier mod Z SET3 PF and cable CAB TS6 supply by Logic The color code of the CAB TS6 cable remains to Table 4 Figure 13 Load cells connection 6 wires with Zener barrier Z SET3 PF 3 5 3 Load cells connection in ATEX Zone 1 active barrier In case of use of an active barrier below is an example of conne...

Page 23: ...tmospheres for connecting up to 4 load cells and go out with a single CAB TS6 extension cable The connections are shown in Table 5 and in Figure 15 Table 5 Connection to the junction box JBOX V3 EX1 Table of signal and color code cable Extension cable CAB TS6 Terminal TB5 Signale Description Color Code 1 GND Shld cable shield GND Shld 2 Exc excitation Red 3 S sense Green 4 Exc excitation Blue 5 S ...

Page 24: ...Db docx Mod L0006A01 of 30 05 2016 Page 24 of 78 Figure 15 Connection to the junction box JBOX V3 EX1 1 TB2 TB1 4 2 1 3 5 4 2 1 3 5 Load Cell 2 Load Cell 1 sig sig Exc S Exc Shld S TB4 TB3 4 TB5 2 3 6 5 7 4 2 1 3 5 4 2 1 3 5 CAB TS6 Extension Cable Load Cell 4 Load Cell 3 ...

Page 25: ...LC3 RD Through the remote display you can set the calibration line weight perform calibration of zero and full scale and edit the configuration of the instrument Refer to Chapter 3 for the operating instructions for use of remote display 3 6 2 RS 485 connection with Logic display DLC3 RD The connection to the remote display Logic DLC3 RD is through an RS 485 serial cable with two wires with the gr...

Page 26: ... bits Stop bit 1 Parity None 3 6 3 RS 485 connection with display BEKA BA488C The connection with the remote display BEKA BA488C in hazardous area must be through the BEKA BA201 barrier as shown in Table 8 and in Figure 17 The settings of the remote display to be set for correct operation are shown in Figure 17 The shield must be connected at one point only cabinet side DLC NET for better immunity...

Page 27: ...eters for communication with DLC NET The parameters also shown in Figure 17 are to be set on BA488C to have a correct communication Table 9 Serial port parameters for communication with BEKA BA488C Parameter Value Transmission Asynchronous Baud rate 19200 Word 8 bits Stop bit 1 Parity None Set up the KeyPad parameter to Internal to use the integrated buttons of BEKA ...

Page 28: ...mmunication port Fieldbus Profibus DP The connection should be carried out on the TB2 connector as shown in the following table Table 10 Table connections to Profibus DP Fieldbus Pin connector Profibus DP on DLC NET Terminal connector Profibus DP Description Wire TB2 3 A Tx Rx Green TB2 8 B Tx Rx Red Figure 18 Instrument connection to Profibus DP Fieldbus Note how the termination on the connector ...

Page 29: ... connector integrated into the instrument will present two RJ45 Ethernet ports allowing cascade connection of devices The connection should be carried out on the TB2 connector as shown in the following table Table 11 Table connections to Profinet Fieldbus Pin connector RJ45 on DLC NET Pin connector RJ45 bus Description Wire TB2 1 1 Tx White Orange TB2 2 2 Tx Orange TB2 3 3 Rx White Green TB2 6 6 R...

Page 30: ...ons by potential free contacts The second typology is shown in the following paragraph The pins have in both cases the functions described in Table 12 Table 12 Table external buttons connections Pin on DLC NET Description TB3 4 0 Vdc TB3 5 24 Vdc TB3 6 TB3 9 Buttons inputs Figure 20 Connection to external buttons through potential free contacts 3 8 2 Inputs from active source In case to interface ...

Page 31: ...he setpoint chosen in the menu and at the intervention mode chosen high or low in Connections are described in Table 14 and Figure 22 It is necessary the expansion board EB Table 14 Connection to setpoint output Pin on DLC NET Description TB6 1 0 Vdc TB6 2 Out 1 24 Vdc transistor open collector PNP setpoint 1 TB6 10 Out 2 24 Vdc transistor open collector PNP setpoint 2 TB6 3 Out 3 24 Vdc transisto...

Page 32: ...stor open collector PNP solenoid valve output instant ingredient 3 TB6 11 Out 4 24 Vdc transistor open collector PNP solenoid valve output instant ingredient 4 TB6 4 Out 5 24 Vdc transistor open collector PNP pump output delayed ingredient 1 TB6 12 Out 6 24 Vdc transistor open collector PNP pump output delayed ingredient 2 TB6 5 Out 7 24 Vdc transistor open collector PNP pump output delayed ingred...

Page 33: ...Series Installation LOGIC S r l M0145Db docx Mod L0006A01 of 30 05 2016 Page 33 of 78 Table 16 Analog output connection Pin on DLC NET Description TB5 4 Schield SHLD TB5 5 I TB5 6 I Figure 24 Analog output connection 0 20 mA or 4 20 mA ...

Page 34: ...set on the printer must be that of Table 18 It is necessary the expansion board EB Table 17 RS 232 connection to the printer Pin on DLC NET Pin on printer TB5 1 GND TB5 2 RX instrument TXD printer TB3 3 TX instrument RXD printer Figure 25 Printer connection through RS232 The RS232 serial port of the printer must be configured as follows in order to be compatible with the DLC NET Table 18 Serial po...

Page 35: ...icates that the corresponding bit is 0 The open output indicates that the corresponding bit is 1 The data weight is updated when required through the BYTE SEL input The most significant byte bit 8 bit 15 it is available valid data after three milliseconds from the input command BYTE SEL Removing the BYTE SEL command the least significant byte bit 0 bit 7 is available after three milliseconds A fur...

Page 36: ...e 19 Connection for parallel interface type PNP Pin on DLC NET Description TB6 1 0 Vdc TB6 2 Out 1 24 Vdc TB6 10 Out 2 24 Vdc TB6 3 Out 3 24 Vdc TB6 11 Out 4 24 Vdc TB6 4 Out 5 24 Vdc TB6 12 Out 6 24 Vdc TB6 5 Out 7 24 Vdc TB6 13 Out 8 24 Vdc TB6 15 Byte Select input 24Vdc Figure 27 Connection for parallel interface type PNP 3 10 3 Connection for parallel interface type NPN The connection of the o...

Page 37: ...onnection for parallel interface type NPN Pin on DLC NET Description TB6 1 0 Vdc TB6 2 Out 1 24 Vdc TB6 10 Out 2 24 Vdc TB6 3 Out 3 24 Vdc TB6 11 Out 4 24 Vdc TB6 4 Out 5 24 Vdc TB6 12 Out 6 24 Vdc TB6 5 Out 7 24 Vdc TB6 13 Out 8 24 Vdc TB6 8 0 Vdc TB6 15 Byte Select input 24Vdc Figure 28 Connection for parallel interface type NPN ...

Page 38: ...USE AND MAINTENANCE MANUAL DLCNET Series Installation LOGIC S r l M0145Db docx Mod L0006A01 of 30 05 2016 Page 38 of 78 INTENTIONALLY BLANK PAGE ...

Page 39: ... to current classifications as described in paragraph 3 5 4 1 2 Setting speed ADC The speed of the analog digital converter is settable to the value of 10 or 80 conversions per second through the switches SW2 1 Figure 29 Switch SW2 1 for setting speed ADC Table 21 Setting speed ADC SW2 1 Description ON 10 samples s OFF 80 samples s The other DIP switch of SW1 must remain at 0 and never be changed ...

Page 40: ...lish communication Figure 31 shows an example of hardware configuration for Siemens PLC in Simatic Step 7 development environment Figure 31 Hardware setting for Siemens PLC For the supply of the GSD library files for Profibus DP Protocol and GSDML for Profinet contact the customer service Logic 4 2 2 Data structure Looking at the PLC side the data structure provides 2 output bytes and 8 bytes of i...

Page 41: ...e watchdog stops to commute purpose to prevent the connected PLC to start dosing during possible phases of the updating of the instrument parameters Weight stability threshold Type Byte Bit Description IN 6 7 Reading weight stability threshold correctly set default 0 The weight stability threshold is an integer value positive 16bit in engineering units that provides the DLC NET information to set ...

Page 42: ...ormation on weight stability and error ADC and the least significant LSB PEB 261 that contains the overflow ADC bit and watchdog In the example all bits are 0 except the watchdog equal to 1 The last word read is the PEW 262 equal to the stable weight threshold set in this case by the word AW 256 4 2 3 Diagnostics On the rear side of the instrument there is LED F for diagnostics of fieldbus communi...

Page 43: ...board the DLC NET as described in paragraph 3 6 In order to ensure the displaying of weight it is necessary to perform the operations of setting and calibration in kg Via Fieldbus the value read by the load cells is always transmitted expressed in divisions 5 1 1 User interface The instrument operator interface is that shown in Figure 33 Figure 33 Operator interface of DLC NET instrument Using the...

Page 44: ...to get in the various steps of parameterization as described in the parameter procedures To shift between different menus In DLC3 RD the printing function is not present During the weight display captures the current weight put zero on the display During the weight display allows the reading of the gross weight and net weight Button used to start a new dosing external button Leds have instead the ...

Page 45: ...ait about 15 minutes for thermal stabilization Will be require a password Enter and press to confirm The instrument will show the written Place the calibrator to 0 mV V Wait no less than 10 seconds and press the button to confirm The instrument will show the written Place the calibrator to 2 mV V Wait no less than 10 seconds and press the button to confirm Will be shown the written and calibration...

Page 46: ...sion board EB When the desired menu is selected press the button to confirm and enter For the menu you must press simultaneously and Press at any time to cancel and return to the weight display screen 5 3 3 Net gross weight displaying On the main screen the system shown the actual weight value By pressing the button you can switch the display from gross weight to net weight and vice versa Function...

Page 47: ...on with sample weights To be made when it has easily sample weights applicable to the weighing system low capacity scales to obtain maximum accuracy Calibration with load cells simulator An alternative to the previous way To be made with high capacity tanks and scales when the load cell sensitivity values are known Pending Accuracy 0 1 Automatic calibration An alternative to the two previous way T...

Page 48: ... start editing the data Display shows the full scale just set Use the button to select the decimal point position and press to confirm and go to the next parameter The data later be requested is the value of zero scale Is then displayed the message In the case of using a simulator for calibration of load cells the user must bring it on the indication 0 mV V In the case of calibration with sample w...

Page 49: ... and press to enter the menu Press at any time to cancel and return to the previous menu The display shows the message because the system waits as first data the full scale of the weighing system Press to start editing the data It displays the full scale currently set with the first digit flashing Use the button to increase the digit and to move on to editing the next one Press to confirm At this ...

Page 50: ...ng the next one Press to confirm The display shows the message the value that the instrument should be displayed instead of the entered in Use the buttons and to increase the digit and move to the next Press to confirm At this point the instrument will show the written and the procedure is finished 5 4 6 Zero scale calibration The procedure can be performed with the following steps From the main s...

Page 51: ... indication of the weight were to decrease the analog output not proportionally follow the indication of the display in the negative but will remain locked to zero To calibrate the analog output it is necessary the following equipment 24Vdc power supply to power the instrument Multimeter for current reading to be connected to the output of DLC NET We recommend using a high precision multimeter ver...

Page 52: ...he display shows the message as the instrument waits for the value of the DAC divisions corresponding to 20mA Press to start editing edit the value using the buttons and until the multimeter does not indicate 20mA default Press to confirm The last parameter to edit is if the analogue output must follow the trend of the net or gross weight The display shows the message corresponding to the current ...

Page 53: ...rating mode Enter the menu as described in paragraph 5 3 2 Scroll through the button until the menu Press to enter Press at any time to cancel and return to the previous menu The display will show the actual operating mode To change it use the button until the desired item is not displayed Press to confirm At this point the instrument will show the letters and the procedure is finished 5 5 3 Confi...

Page 54: ...play and thermal printer For the instrument equipped with EB expansion board are available three different combinations printer RS 232 and remote display DLC3 RD RS 485 RS 232 printer and remote display BEKA BA488C RS 485 remote display BEKA BA488C RS 232 and RS 485 printer Viceversa without EB expansion board the only RS 485 port is available and two configuration options remote display DLC3 RD R...

Page 55: ...currently set default To change it use the button until the desired item is not displayed Press to confirm At this point the instrument will show the letters and the procedure is finished 5 5 7 Minimum increase configuration The parameter is the minimum increase in engineering units which is used to change the weight on the display To change the parameter Enter the menu as described in paragraph 5...

Page 56: ... message with the value currently set default To change it use the button until the desired item is not displayed Press to confirm At this point the instrument will show the letters and the procedure is finished 5 5 9 Real time clock configuration On the EB expansion board there is a real time clock for clock calendar storage for printing To chance and update the real time clock follow this proced...

Page 57: ...eight of the scale This means that the tare button does not change the set value It is possible to select for each of setpoint the way of action refers to the net weight In this mode when you press the button the set point is recalculated automatically and stored that is added in algebraic mode to the set the tare value previously acquired The setpoint concerned shall take both increase that to de...

Page 58: ...high alarm or low set point 2 work on the net weight high alarm or low set point 3 work on the gross weight high alarm or low set point 3 work on the net weight high alarm or low set point 4 work on the gross weight high alarm or low set point 4 work on the net weight high alarm or low When you arrive on the fourth set point and pressing the editations are saved 5 6 2 Setpoint operating value To s...

Page 59: ...USE AND MAINTENANCE MANUAL DLCNET Series Use as interface with indicator LOGIC S r l M0145Db docx Mod L0006A01 of 30 05 2016 Page 59 of 78 Press to exit and return to the previous menu ...

Page 60: ...USE AND MAINTENANCE MANUAL DLCNET Series Use as interface with indicator LOGIC S r l M0145Db docx Mod L0006A01 of 30 05 2016 Page 60 of 78 BLANK PAGE ...

Page 61: ... with EB expansion board are equipped with 8 outputs and 1 additional input for dosing up to 4 components For each component two outputs are provided one for the solenoid valve and one for the pump To use the DLC NET as a doser set the parameter of the menu as by following the instructions in paragraph 5 5 2 6 1 2 Dosing control The application of the system is that shown in Figure 36 Figure 36 Ap...

Page 62: ...ks the necessary conditions for the beginning of dosing If the agreement is positive automatic self tare is made and the corresponding valve output is activated eg ingredient 1 OUT 1 After 2000 ms the output corresponding to pump is activated eg ingredient 1 OUT 5 Both outputs remain active until the scale indicates a net weight equal to the setpoint minus the amount of in flight and pre action Ne...

Page 63: ...start the dosage In other words before starting a dosing the scale must score a gross weight equal to least the amount set in parameter This parameter can be used as protection in case of absence of the tank tin before dosing To set the zero scale Enter the menu as described in paragraph 5 3 2 Scroll through the button until the menu Press to enter Scroll through the button until the menu Press to...

Page 64: ...previous menu The display will show the value currently set default To change it use the button to increase the digit and to move on to editing the next one Press to confirm At this point the instrument will show the letters and the procedure is finished If the parameter is set to 0 it is not performed any capacity control and are allowed dosages that could lead to spills in the event of incorrect...

Page 65: ...e time in seconds that the instrument waits after closing the automatic dosing valve before acquiring the data weight for printing In other words when the valve closes when it was reached a net weight equal to the setpoint minus the in flight amount the instrument considers the data stable after seconds For an accurate reporting choose this parameter in sufficiently large way to be sure that the a...

Page 66: ...Instrument ID Start time of printing First automatic manual dosing component and actual value Last automatic manual dosing component and actual value the recipe in print mode up to 14 components is active If you print before finishing the recipe through command see par 6 4 4 is printed the report component used up to that point The end time will be empty 0000 00 00 00 00 If you release after the c...

Page 67: ...the main weight display screen press the button to enter the top level menu Use the button to scroll through the menu items to the section Press to enter the menu Press at any time to cancel and return to the previous menu The display will show as a first option equal to the queue for the ingredient 1 Press to edit the first queue press to cycle instead edit the next queue Entering the edit menu o...

Page 68: ...me to cancel and return to the previous menu The display will show as a first option equal to the queue for the ingredient 1 Press to scroll to up to the first of the operations Press to choose the pre actions to edit When it choose press to confirm and start to edit Entering the edit menu of the single pre actions the display will show with the first digit flashing Use the button to increase the ...

Page 69: ...inting Once the entire sequence is complete you can finish the recipe and print the report of the whole process including start time and end time recipe 6 4 2 Dosing of automatic components To dose the automatic components follow these steps Press the button to enter the ingredients selection menu The display will show the written corresponding to the first automatic ingredient Use the button to s...

Page 70: ...minus the in flight amount the system waits for the waiting time of the in flight set amount and shows the written acquiring the data weight for the printer At this point the led corresponding to the component dosing is steady on indicating that the filling process is ended The led continues to blink and the dosage is terminated To dose a new automatic component repeat the full procedure To dose a...

Page 71: ...of pressing the button the display will show and returns automatically to the main menu After the dosage of one or more manual components the led is steady on indicating that the last dosage of manual components has been successfully 6 4 4 End of recipe If a recipe is in progress you can end it at any time with command This operation assigns the end time recipe Once done the you can print the last...

Page 72: ...rmed the print will be sent automatically via the RS 232 serial port 6 4 6 Abort of recipe If there is any malfunction it is possible to abort a recipe To abort a recipe follow the procedure below Press the buttons and simultaneously The display will show The printer memory is erased all the leds are turned off and you can start a new recipe If you make abort during a dosage of automatic component...

Page 73: ... work only with staff specialized in this type of operation If necessary contact the organizations specialized in this kind of activity 7 2 Procedure The instrument does not have in its component parts defined toxic or harmful The instrument consists essentially of the following materials Ferrous metal material Non ferrous metal material Electrical components wiring and circuit boards The dismantl...

Page 74: ...USE AND MAINTENANCE MANUAL DLCNET Series Decommissioning and dismantling LOGIC S r l M0145Db docx Mod L0006A01 of 30 05 2016 Page 74 of 78 INTENTIONALLY BLANK PAGE ...

Page 75: ...on of DLC NET to display BEKA BA488C 26 Figure 18 Instrument connection to Profibus DP Fieldbus 28 Figure 19 Instrument connection to Profinet Fieldbus 29 Figure 20 Connection to external buttons through potential free contacts 30 Figure 21 External buttons connection from active source 31 Figure 22 Connection to setpoint output 31 Figure 23 Input and output connections for dosing 32 Figure 24 Ana...

Page 76: ...USE AND MAINTENANCE MANUAL DLCNET Series FIGURES LOGIC S r l M0145Db docx Mod L0006A01 of 30 05 2016 Page 76 of 78 INTENTIONALLY BLANK PAGE ...

Page 77: ...munication with BEKA BA488C 27 Table 10 Table connections to Profibus DP Fieldbus 28 Table 11 Table connections to Profinet Fieldbus 29 Table 12 Table external buttons connections 30 Table 13 Table connections from active source 30 Table 14 Connection to setpoint output 31 Table 15 I O connections in Batch mode 32 Table 16 Analog output connection 33 Table 17 RS 232 connection to the printer 34 Ta...

Page 78: ...USE AND MAINTENANCE MANUAL DLCNET Series TABLES LOGIC S r l M0145Db docx Mod L0006A01 of 30 05 2016 Page 78 of 78 INTENTIONALLY BLANK PAGE ...

Reviews: