Using the IntelliLogger™
35
Figure 3-9; Integrated ILIM-7 Configuration Switches and Wiring Connections
A
NALOG
I
NPUTS
Eight bipolar analog inputs are provided on the ILIM-7 and are identified as Channels A
through H. All eight channels are identical in range and input type offering with the exception
that the first four channels (A through D) offer an additional range (Vdc-Hi) that allows for
direct connection of Vdc signals up to +/-30Vdc.
Analog input channels are configured for current or voltage/thermocouple via a DIP switch
setting on a channel by channel basis. Within each input type, multiple full-scale ranges are
offered and specified via the HyperWare-II Program Net setup.
Input Types
The following inputs can be connected to the ILIM-7. Input ranges are specified
in Appendix D (ILIM-7 Module Icon Reference) in the HyperWare-II manual.