2.3.3 NG CASTORE XS iQ –
FULL RANGE power supply voltage
Models: NG CASTORE XS iQ
18
24
36
N2 outlet
Flow rate (Max)
18 l/min
24 l/min
36 l/min
Outlet pressure (Max)
8 Bar (116 psi)
Nitrogen purity
LCMS grade (up to 99.9%
*1
)
Dew point
*2
< -60°C (-76°F)
Communication
LCD with touch screen
Standard
RS485
Standard
RS232
For service
General data
Power supply voltage (min-Max)
115-240 Vac (±10%) 50/60 Hz
Connection type
IEC C20
Nominal power (Max)
1.2 KW
1.3 KW
1.5 KW
Main fuse F1(10x38mm)
15A (600Vac – T – class CC)
Inverter protection fuses F3 - F4
(10x38mm)
15A (300Vdc – T – class CC)
Main-board fuse F5 (5x20mm)
3.15A (250Vac – T – class CC)
Net weight
145 kg
Noise level
< 48 dB
< 50 dB
Heat value (BTU)
4100
4440
5130
Dimensions (W x D x H)
51 x 85 x 69 cm
Connections
N2 outlet port(s)
¼" BSPP female
Drain port
¼" BSPP female
*1
The purity refers to the residual oxygen, depends on flow and working pressure
*2
Atmospheric Dew Point (ADP)
Table 3: NG CASTORE XS iQ – Full range power supply voltage