LMI 585 Plus Mastering Instructions Download Page 3

E-mail support 

[email protected]

 

LMI Corporation

Phone support 

(810) 714-5811 

 

Form:  CA 068 

August 01, 2005 

R:\Quality\Calibration Instructions\CA 097 200s or sb to LMI 585 Plus.doc 

Page 3 of  4 

 

 

 

¾

 

CAL LOW will be displayed on 
the screen.  Insert the 200 probe in 
the “LO” step of the 210 Master 
block and press <SELECT> 

     

 

 
 
 
 
 

 

 

 

 

 

 

 

 

 

      

¾

 

Calibration is complete and the screen takes you back to the “Calibrate Menu:” screen.  To exit 
and view your readings, press <SEND>. 

 

 
 
 
 
 
 

¾

 

You should be able to view your readings on the screen labeled 
1PROBE.  Use the 

 or 

 to toggle between three gage screens 

if 1PROBE is not shown.  The screen below shows 1PROBE with 
a gage reading of 0.70mm.  Insert the gage back into the “31.0 
Master” step to verify zero or master.

 

This calibration procedure 

will produce a negative reading when the Probe is extended 
beyond the 31.0 Master Point.  

 

¾

 

The screen will then toggle to 
CAL HIGH.  Insert the 200 
probe in the “HI” step of the 
210 Master block and press 
<SELECT>

 

¾

 

The screen will then toggle to 
CAL MASTER.  Insert the 200 
probe in the “31.0 Master” step 
of the 210 Master block and 
press <SELECT>

 