Form: CA 068
August 01, 2005
R:\Quality\Calibration Instructions\CA 097 200s or sb to LMI 585 Plus.doc
Page 3 of 4
¾
CAL LOW will be displayed on
the screen. Insert the 200 probe in
the “LO” step of the 210 Master
block and press <SELECT>
¾
Calibration is complete and the screen takes you back to the “Calibrate Menu:” screen. To exit
and view your readings, press <SEND>.
¾
You should be able to view your readings on the screen labeled
1PROBE. Use the
►
or
◄
to toggle between three gage screens
if 1PROBE is not shown. The screen below shows 1PROBE with
a gage reading of 0.70mm. Insert the gage back into the “31.0
Master” step to verify zero or master.
This calibration procedure
will produce a negative reading when the Probe is extended
beyond the 31.0 Master Point.
¾
The screen will then toggle to
CAL HIGH. Insert the 200
probe in the “HI” step of the
210 Master block and press
<SELECT>
¾
The screen will then toggle to
CAL MASTER. Insert the 200
probe in the “31.0 Master” step
of the 210 Master block and
press <SELECT>