LMI 585 Plus Mastering Instructions Download Page 2

E-mail support 

[email protected]

 

LMI Corporation

Phone support 

(810) 714-5811 

 

Form:  CA 068 

August 01, 2005 

R:\Quality\Calibration Instructions\CA 097 200s or sb to LMI 585 Plus.doc 

Page 2 of  4 

 

 

 

*NOTE* 

These instructions are copied from the LMI 585 

Plus

 User Guide and are intended to be used as a 

quick reference for the mastering process. Please refer to the LMI 585 

Plus

 User Guide for the setup 

and configuration of the LMI gage. 

 
 
Prior to starting calibration, make sure your 200 Probe is connected to the Gage Port #1  on the 585 
Plus using a 6009 cable.   
 
 
 
 
 
 
 
 
Under the ADVANCED SETUP/CONFIG PORTS menu, the CALSPAN of Port 1a should be set to 
10.00mm. Reference LMI 585 Plus User Guide section 6 sub section 6.2.2 
 

¾

 

Press <MENU> to enter the main menu 

 

¾

 

“Calibrate” should already be highlighted like the figure below.  Press <SELECT> or 

 

 
 
 
 
 
 
 
 

 

¾

 

Highlight Port 1a then press <SELECT>  

 
 
 
 
 
 
 
 

 
 
 
 