14 of 92 — 26-0008-1100/5
(Worldwide)
Physician’s Manual — VNS Therapy Programming Software (8.1)
4. I
NTERROGATE
D
EVICE
________________
To retrieve information from the Pulse Generator or to program the Pulse
Generator, you must first select the
Interrogate Device
button from the
MAIN Menu. The START INTERROGATION Screen will be displayed
(see Figure 2).
Figure 2.
START INTERROGATION Screen
To interrogate the Pulse Generator, place the Programming Wand over the
Pulse Generator and tap
Start Interrogation
. Tap
Cancel
to return to the
previous screen.
Upon completion of a successful interrogation, the Programming Software
automatically displays the PARAMETER screen (see Figure 3 on page 15)
unless an error condition (i.e., Generator End of Service or High Lead
Impedance) is detected with the Pulse Generator. If an error condition is
detected with the Pulse Generator, the associated warning message is
displayed and must be acknowledged by the user before proceeding to the
PARAMETER screen. The PARAMETER screen displays the operating
parameters of the Pulse Generator with their current settings.
There is one line for each of the eight parameters. Each line contains three
types of information:
1. The name of the parameter and the units used to express its values
2. The present setting for that particular parameter (***** is shown for
any unknown or out of range parameter values)