Operating conditions:
Temperature
Relative humidity
5ºС ~ 40ºС
<80%
Storage and transportation environment:
Temperature
Relative humidity
-20 ºС ~ 55ºС
< 93%
Net Weight (without batteries)
31g
External dimension
58 (L) х 30 (W) х 34 (H) mm.
Date of manufacturing
Manufacturing Date printed on sticker
Country of Origin
PRC
Symbol Defi nitions:
SYMBOL
DEFINITION
SYMBOL
DEFINITION
Type BF applied part.
Conformity to WEEE Directive
IP22
Protected against dripping water.
Attention
PR bpm
Pulse rate (BPM)
%SpO2
Oxygen saturation
No SpO2 Alarm
Low power indication
Storage temperature and relative
humidity
Serial No.
Date of Manufacture
Follow instructions for use
European Union approval
Authorized representative in the
European community
Sign of type approval of measuring
instruments
Manufacturer’s information
DECLARATION
This device complies with EMC (IEC 60601-1-2:2001, CISPR 11/A2:2002
(Group 1, Class B), IEC 61000-4-2:2001, IEC 61000-4-3:2002, IEC 61000-4-8:2001).
The materials which user can come into contact have no toxicity and no action on tissues, complying with ISO10993-
1,-5,-10.
Manufactured under control and for Little Doctor International (S) Pte. Ltd., 7500A, Beach Road, 11-313 The Plaza
199591, Singapore. Yishun Central P.O. Box 9293 Singapore 917699.
Manufacturer: Beijing Choice Electronic Technology Co., Ltd. (Room 4104, №.A12 Yuquan Road, Haidian District,
100143 Beijing, PRC).
Distributor in Europe: Little Doctor Europe Sp. z o.o., 57G Zawila Street, 30-390, Kraków, Poland
Sales Offi
ce phone: +48 12 2684746, 12 2684747, fax: +48 12 268 47 53
E-mail: [email protected]
www.LittleDoctor.pl
GUARANTEE
We grant 2 years guarantee on the product commencing on the date purchase. The guarantee only comes info force
if the data of purchase is confi rmed by the dealer’s stamp and signature on Guarantee Card.
This warranty does not extend to, and will be void in respect of any products which have been subjected to misuse,
neglect, fi re, improper modifi cation, use in violation of the instructions furnished by us repaired by an unauthorized
third party.
TROUBLESHOUTING
Problems
Possible reason
Solution
SpO
2
or
PR can not
be shown
normally
1. Finger is not plugged
correctly
2. Patient’s Oxyhemoglobin
value is too low to be measured
1. Retry by plugging the fi nger
2. Measure other patients to make sure that no problem
exists in the product. Go to a hospital in a timely manner
for an exact diagnosis.
SpO
2
or PR
is shown
unstably
1. Finger might not be plugged
deep enough
2. Finger is trembling or
patient’s body is in movement
status
1. Retry by plugging the fi nger
2. Try not to move
The Oximeter
can not be
powered on
1. Power of batteries might be
inadequate or not be there
at all
2. Batteries might be installed
incorrectly
3. The Oximeter might be
damaged
1. Please replace batteries
2. Please reinstall the batteries
3. Please contact with local customer service centre
Indication
lamps are
suddenly off
1. The product is automatically
powered off when no signal is
detected longer than 8 seconds
2. Power is too low.
1. Normal
2. Replace the batteries
E595/2111/4_ENG
MD300C23
Name and address of retailer
Serial №
MODEL
Date of Purchase
Purchaser sign
I confi rm that surface appearance
and complete set of device is OK:
Retailer stamp
Purchaser sign
GUARANTEE CARD
For more information please visit
WWW.LITTLEDOCTOR.SG
LITTLE DOCTOR INTERNATIONAL (S) PTE. LTD.
Yishun Central P.O. Box 9293 Singapore 917699,
Fax: 65-62342197, E-mail: [email protected]
EC-Rep: Shanghai International Holding Gorp. GmbH (Europe).
Eiff estrasse 80, 20537 Hamburg, Germany.
Beijing Choice Electronic Technology Co., Ltd.,
Room 4104, №.A12 Yuquan Road, Haidian District,
100143 Beijing, PRC
“Error3” or
“Error4”
Displayed on
screen
1 .Low power
2. Receiving tube being
shielded or damaged together
with broken connector.
3. Mechanical Misplace for
receive-emission tube
4. Amp circuit is malfunction.
1. Change new battery
2. Please contact with local customer service center
3. Please contact with local customer service center
4 Please contact with local customer service center
“Error7”
displayed on
screen
1. Low power
2. Emission tube damaged.
3. Current control circuit is
malfunction.
1 Change battery
2 Contact with local customer service center
3 Contact with local customer service center
ELECTROMAGNETIC COMPATIBILITY
The device conforms to IEC60601-1-2:2014 Electromagnetic Compatibility (EMC) standard.
Essential performance is defi ned as SpO2 accuracy and pulse rate accuracy or an indication of abnormal operation.
Accuracies may be aff ected as a result of exposure to electromagnetic disturbances that are outside of the environ-
ments listed in the intended use. If issues are experienced, move the device away from the source of electromagne-
tic disturbances.
TABLE 1: ELECTROMAGNETIC EMISSIONS LIMITS AND COMPLIANCE
Emissions Test
Compliance
RF Emissions
CISPR 11
Group 1, Class B
Note: Harmonic Emissions (IEC 61000-3-2), Voltage Flicker Emissions (IEC 61000-3-3) are not applicable.
TABLE 2: ELECTROMAGNETIC IMMUNITY
Immunity Test
Compliance
Electrostatic Discharge (ESD)
IEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
Rated power Frequency Magnetic Fields
IEC 61000-4-8
30 A/m
50Hz and 60 Hz
Radiated RF
IEC 61000-4-3
80 MHz – 2.7 GHz
10 V/m 80% AM 1kHz
380 – 390 MHz
27 V/m Pulse mod. 18Hz
430 – 470 MHz
28 V/m FM±5Hz deviation 1kHz sine
704 – 787 MHz
9 V/m Pulse mod. 217Hz
800 – 960 MHz
28 V/m Pulse mod. 18Hz
1.7 – 1.99 GHz
28 V/m Pulse mod. 217Hz
2.4 – 2.57 GHz
28 V/m Pulse mod. 217Hz
5.1 – 5.8 GHz
9 V/m Pulse mod. 217Hz
Note: Electrical Fast Transients (IEC 61000-4-4), Surge (IEC 61000-4-5), Voltage dips (IEC 61000-4-11),
Conducted Immunity (IEC 61000-4-6) are not applicable.