background image

INSTRUCTION MANUAL

Fingertip Pulse Oximeter model MD300C23

GENERAL DESCRIPTION

Oxygen Saturation is a percentage of Oxyhemoglobin (HbO2) capacity, compounded with oxygen, by all combinative 
hemoglobin (Hb) capacity in blood. In other words, it is consistency of Oxyhemoglobin in blood. It is a very important 
parameter for the Respiratory circulation System. 

Many respiratory diseases can result in oxygen saturation being lowered in human blood. Additionally, the following 
factors can reduce oxygen saturation: Automatic regulation of organ dysfunction caused by Anesthesia, Intensive Posto-
perative Trauma, injuries caused by some medical examinations. 

That situation might result in light-headedness, asthenia, and vomiting. Therefore, it is very important to know the oxy-
gen saturation of a patient so that doctors can fi nd problems in a timely  manner.

The fi ngertip pulse Oximeter features small size, low power consumption, convenient operation and portability. 
It is only necessary for a patient to put one of his fi ngers into the fi ngertip photoelectric sensor for diagnosis, and 
a display screen will show oxygen saturation. 

PARTS LIST

Complete set (fi g. 1): 
Main Unit – 1 pcs., 
Instruction Manual with Warranty Card – 1 pcs.,  
Hang Lace – 1 pcs., 
batteries – 2 pcs. 

MEASUREMENT PRINCIPLE

Principle of the Oximeter is as follows: A mathematical formula is established making use of Lambert Beer Law 
according to Spectrum Absorption Characteristics of Reductive hemoglobin (RHb) and Oxyhemoglobin (HbO2) in 
glow and near-infrared zones. 

Operation principle of the instrument is Photoelectric Oxyhemoglobin Inspection Technology is adopted in accor-
dance with Capacity Pulse Scanning and Recording Technology, so that two beams of diff erent wavelength of lights 
(660nm glow and 940nm near infrared light) can be focused onto a human nail tip through a clamping fi nger-type 
sensor. 

A  measured  signal  obtained  by  a  photosensitive  element,  will  be  shown  on  OLED 
display through process in electronic circuits and microprocessor.

Diagram of Operation Principle                                                                                                         

 1.  1.Red and Infrared-ray Receipt Tube.                                

 2.  1.Red and Infrared-ray Emission Tube. (fi g.2)  

INTENDED USE

The Fingertip Pulse Oximeter is a handheld non-invasive device intended for spot-checking of oxygen saturation 
of arterial hemoglobin (SpO2) and Pulse Rate of adult, adolescent and child patients in hospitals, hospital-type 
facilities and homecare.

PRECAUTIONS FOR USE

1.  Before use, carefully read the manual.

2.  Operation of the pulse oximeter may be aff ected by the use of an electrosurgical unit (ESU).

3.  The pulse oximeter must be able to measure the pulse properly to obtain an accurate SpO2 measurement. Verify 

that nothing is hindering the pulse measurement before relying on the SpO2 measurement.

4.  Do not use the pulse oximeter in an MRI or CT environment.

5.  Do not use the pulse oximeter in situations where alarms are required. The device has no alarms. It is not for 

continuous monitoring.

6.  Do not use the pulse oximeter in an explosive atmosphere.

7.  The pulse oximeter is intended only as an adjunct in patient assessment. It must be used in conjunction with 

other methods of assessing clinical signs and symptoms.

8.  In order to ensure correct sensor alignment and skin integrity, the maximum application time at a single site for 

our device should be less than half an hour.

9.  Do not sterilize the device using autoclaving, ethylene oxide sterilizing, or immersing the device in liquid. The 

device is not intended for sterilization.

10. Follow local ordinances and recycling instructions regarding disposal or recycling of the device and device com-

ponents, including batteries.

11. This equipment complies with IEC 60601-1-2:2014 for electromagnetic compatibility for medical electrical equ-

ipment and/or systems. However, because of the proliferation of radio-frequency transmitting equipment and 
other sources of electrical noise in healthcare and other environments, it is possible that high levels of such 
interference due to close proximity or strength of a source might disrupt the performance of this device.

12. Portable and mobile RF communications equipment can aff ect medical electrical equipment. The portable and 

mobile RF communications equipment should be used no closer than 30cm (12 inches) to any part of the device, 
including cables specifi ed by the manufacturer. Otherwise, degradation of the performance of this equipment 
could result. 

13. This equipment is not intended for use during patient transport outside the healthcare facility. 

14. The patient is an intended operator. All functions of the device can be safely used.

15. It may be unsafe to:

  –  use accessories, detachable parts and materials not described in the instructions for use 

  –  interconnect this equipment with other equipment not described in the instructions for use

  –  disassemble, repair or modify the equipment

16. The material that contact with the patient’s skin has passed the ISO10993-5 Tests for invitro cytotoxicity and 

ISO10993-10 Tests for irritation and delayed-type hypersensitivity.

17. Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result 

in improper operation. If such use is necessary, this equipment and the other equipment should be observed to 
verify that they are operating normally.

18. The use of accessories, transducers and cables other than those specifi ed or provided by the manufacturer of 

this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of 
this equipment and result in improper operation.

19. When the signal is not stable, the reading may be inaccurate. Please do not refer to the measurement. 

20. The material of the device has no nature latex.

21. The pulse oximeter equipment is calibrated to display functional oxygen saturation.

22. The waveform we provide is normalized.

Rx only: „Caution: Federal law (USA) restricts this device to sale by or on the order of a licensed practitioner.”

BATTERY INSTALLATION

 

1.  Shift the bottom panel and then open it (fi g.3a).

2.  Put the two AAA batteries into battery cassette 

in correct polarities (fi g. 3b).

3.  Close the bottom panel (fi g. 3c).

 WARNING!

Replace batteries when low power indncator “ 

 ” is lit or if after pressing Power button   there is no indica-

tion on display. 

Please remove the battery if the Oximeter will not be used for a long time.

Do not use re-chargable batteries.

HOW TO USE

 WARNING!

 Do not tremble the hand with oximeter during measuremrnt

1.  Open the clamp (fi g. 4)
2.  Plug one of fi ngers into rubber hole of the oximeter (it is best to plug the fi nger 

thoroughly) before releasing the clamp.

3.  Press the button   once on front panel.  

DISPLAY MODES.

After turning on the oximeter, each time you press the power switch, the oximeter will switch to another display mode, 
there are 6 display modes shown as follows:

SpO

2%

PR

bpm

95

84

SpO

2%

PR

bpm

97

74

SpO

2%

PR

bpm

98

77

SpO

2%

PR

bpm

98

77

SpO

2%

PR

bpm

97

77

SpO

2%

PR

bpm

97

74

4.  Read correspondent data from display screen:

SpO

2%

PR

bpm

97

74

Low power indicator

Pulse Wave

 

SpO

2%

PR

bpm

95

84

PR signal 
intensity Bar 
graph

PR

SpO

2

BRIGHTNESS ADJUSTMENT

When you press the power switch for a long time (more than one second), the brightness of the oximeter will 
be changed by degrees, there are 10 levels on brightness; the default level is level four. 

HANG LACE INSTALLATION

1.  Thread thinner end of the hang lace through the hanging hole (fi g. 5). 

2.  Thread thicker and of the lace through the threaded end before pulling it tightly. 

MAINTENANCE AND STORAGE

Replace the batteries timely when indncator “ 

 ” is lit.

Clean surface of the fi ngertip oximeter before it is used in diagnosis for patients

Remove the batteries inside the battery cassette if the Oximeter will not be operated for a long time.

It is best to preserve the product in a place where ambient temperatures -20-55r and humidity is <93%.

It is recommended that the product should be kept in a dry environment anytime. A wet ambient might aff ect its 
lifetime and even might damage the product.

Please follow the law of the local govern ment to deal with used battery.

SPECIFICATION

Model

MD300C23

Display

OLED display

SpO

 measurement 

range

 accuracy

70% - 100%
 ±2% 

Pulse Rate
 measurement 

range

 accuracy

30 - 250 bpm
30 - 99 bpm - ±2 bpm;
100 - 250 bpm - ±2% 

Measurement wavelengths
  red   
  infrared

660 nm 
905 nm

Power source

2 x 1,5V AAA (LR03)

Battery set operating period 

> 30 hrs.

Max. power consumption

0.075 W

Applicable fi nger circumference 

20-75 mm

 Fig. 5

 Fig. 4

 Fig. 2

1

2

 Fig. 1

Fig. 3a

 Fig. 3b

 Fig. 3c

ENG

Reviews: