5. On the fi xed cuff  the sign «INDEX» should point to the area  
«NORMAL» (fi g 7). It means that the cuff  is chosen correctly and fi ts 
the size of your upper arm. If the sign points to the area marked «

» or to the left, the cuff  is too small and the readings will be 

higher. If the sign points to the area marked «

» or to the right, 

the cuff  is too large and the readings will be lower.

Fig. 7

6. If the arm has a conic form, the cuff  should be put on 
with a spiral movement as shown in the fi g. 8.

Fig. 8

7. If the rolled-up sleeve squeezes the arm interfering 
with free blood fl ow the Device may give inaccurate fi g-
ures not corresponding to your actual blood pressure 
(fi g. 9).

Fig. 9

8. Place stethoscope head so that it was located in depres-
sion, somewhat higher than elbow bend (Fig. 10).

Fig. 10

9. Pressure shall be measured in sitting or lying position of human. IN SITTING 
POSITION WATCH THAT PART OF HAND WITH CUFF WOULD BE LOCATED AT THE 
LEVEL OF HEART, AND HAND WOULD BE FREELY LOCATED ON THE TABLE AND 
WOULD NOT MOVE.

MEASUREMENT PROCEDURE

1. Insert metal tubes of stethoscope into ears. Close air valve on air pump, 

having turned it clockwise. Squeezing air pump, pump up the cuff , listen-
ing to pulse by stethoscope. After You stopped to listen to pulse, pump 
up the cuff  by 30 mm Hg more.

2. Slowly opening air valve, turning it counterclockwise, bleed down pressure 

in the cuff . Watch that pressure in the cuff  would fall down at speed 2-4 mm 
Hg per second. This is necessary to get accurate result.

3. As soon as You hear weak pulse beats, memorize manometer reading. This 

is Your systolic (upper) arterial pressure.

4. Pressure in the cuff  is continuing falling down at the same speed (2-4 mm Hg 

per second). You are continuing to listen to pulse. Sounds which You hear will 
change. Unlike the fi rst beats, they will become softer, resembling rustling. 
At the moment when You practically stop seize pulse, memorize manometer 
reading. This is Your diastolic (lower) arterial pressure.

Summary of Contents for LD-91

Page 1: ...Ci nieniomierz mechaniczny LD do pomiaru ci nienia t tniczego krwi Instrukcja obs ugi Blood Pressure Monitor Instruction Manual POL ENG LD 91...

Page 2: ...ticoagulants antithrombo cytes or steroids Use of device in these cases may cause internal hemorrhage 2 For correct measurement you should know that THE BLOOD PRESSURE IS SUBJECT TO SHARP VARIATIONS E...

Page 3: ...UR BLOOD PRESSURE WITH THE ELECTRONIC DEVICE Measure your blood pressure in the calm and comfortable conditions at the indoor tem perature No eating an hour before measurement no smoking taking tonic...

Page 4: ...so that the air tube is directed to your palm g 4 If the measurement on your left arm is di cult you may use your right arm In this case remember that the readings may di er by 5 10 mmHg and even more...

Page 5: ...CUFF WOULD BE LOCATED AT THE LEVEL OF HEART AND HAND WOULD BE FREELY LOCATED ON THE TABLE AND WOULD NOT MOVE MEASUREMENT PROCEDURE 1 Insert metal tubes of stethoscope into ears Close air valve on air...

Page 6: ...o time to authorized repair organizations to check the technical condition of the Device 8 Dispose of the Device and its components according to the application local regulations No special requiremen...

Page 7: ...d lower Cu size OF the Device adult s for upper arm circle from 25 to 36 cm Weight without package case g 332 Dimensions package mm 115 x 185 x 75 Year of manufacture Year the manufacture is given in...

Page 8: ...G Zawila Street 30 390 Krak w Poland Sales O ce phone 48 12 2684746 12 2684747 fax 48 12 268 47 53 E mail biuro littledoctor pl www LittleDoctor pl Authorized Representative in the EU Little Doctor Eu...

Page 9: ...ia prawid owego pomiaru nale y pami ta e CI NIENIE T TNICZE PODLEGA SILNYM WAHANIOM NAWET W KR TKIM PRZEDZIALE CZASOWYM Warto ci nienia t tniczego krwi zale y od wielu czynnik w Zwykle jest ona ni sza...

Page 10: ...dnej dla pacjenta w temperaturze pokojowej Na godzin przed pomiarem nie nale y spo ywa posi k w a w ci gu 1 5 2 godzin nie pali nie pi napoj w wzmacniaj cych i alkoholu 5 Dok adno pomiaru ci nienia t...

Page 11: ...j r ce jest utrudnio ny mo na dokona go na prawej Nale y pami ta e wynik mo e si r ni o 5 10 mmHg a w niekt rych przypadkach nawet wi cej Rys 4 3 Owi mankiet tak aby jego dolna kraw d znajdowa a si w...

Page 12: ...CA A R KA LU NO LE A A NA STOLE I NIE PORUSZ A SI PROCEDURA POMIARU 1 Za o y stetoskop na uszy Zamkn zaw r na gruszce przekr caj c go w kierunku zgodnym z ruchem wskaz wek zegara ciskaj c gruszk pompo...

Page 13: ...nktu serwisowego w celu sprawdzenia technicznego stanu urz dzenia 8 Utylizacja urz dzenia przeprowadzana jest wed ug zasad obowi zuj cych obecnie w danym kraju Specjalne warunki utylizacji nie zosta y...

Page 14: ...szka LD S051 1 szt 5 Wlot powietrza LD S056 1 szt 6 Zaw r LD S054 1 szt 7 Etui LD S059 1 szt 8 Instrukcja obs ugi 1 szt 9 Karta gwarancyjna 1 szt 10 Opakowanie 1 szt DOK ADNO POMIARU Produkcja urz dze...

Page 15: ...enia cm Materia Wk adka powietrzna C1N dla noworodk w 7 12 bawe na TPU LD S01N bez pier cienia C1I dla niemowl t 11 19 bawe na TPU LD S01I C1C dla dzieci N1C dla dzieci 18 26 bawe na TPU nylon TPU LD...

Page 16: ...Pte Ltd Little Doctor International S Pte Ltd 2016 LITTLE DOCTOR INTERNATIONAL S PTE LTD Yishun Central P O Box 9293 Singapore 917699 Fax 65 62342197 E mail ld singaporemail com Little Doctor Europe...

Reviews: