22
Important
!
Each patient is encouraged to use personal
mouthpiece, a mask and / or nozzle for the nose.
7. Plug the power cord to an electrical outlet.
8. Connect one end of the inhalation tube to the
fi
tting of com-
pressor, and others - to the
fi
tting of nebulizer.
9. Turn on the nebulizer, switching the power switch in position
«I». NEBULIZER IS READY FOR INHALATION.
Depending on the type of baf
fl
e, the particles of different sizes are dis-
tributed aerosols as follows*:
particle sizes, mkm
% to the total mass
0
5
10
15
1
100
60
30
10
7
5
3
(mainly the upper
respiratory tract)
Dispenser "C" (red)
Dispenser "B" (blue)
Dispenser "A" (yellow)
(universal)
(mainly lower respiratory tract)
For effective delivery medicines to a specific
site of respiratory tract, need to use the
appropriate dispenser.
particle sizes, mkm
% to the total mass
1
100
60
30
10
7
5
3
0
5
10
15
particle sizes, mkm
% to the total mass
1
100
60
30
10
7
5
3
0
5
10
15
Fig. 4
* - Measured by Little Doctor International (S) Pte. Ltd.
Fig. 3