SYMBOL EXPLANATION
BF type product
Important: Read the instruction
Protection class II
Manufacturer
Compliance with the Directive 93/42/EEC
Representative in the EU
The revision date of this Operation manual is indicated at the last page in the EXXX/YYMM/NN format, where YY is
the year, ММ is the month, and NN is the revision number. Technical specifications are subject to change without prior
notice in order to improve operational properties and quality of the product.
TROUBLESHOOTING. DETECTION AND CORRECTION
At occurrence of any problems when using the device, refer to the table below for possible causes of failure.
Problem
Cause
Recommended steps
The intensity of
aerosol formation
is very low
No contact between inhalation
solution and mesh membrane
for more than 10 sec.
Change the angle of inclination of nebulizer so
that the inhalation solution contacts with mesh
membrane.
MESH membrane of inhalation
chamber is clogged.
Use not recommended
Medicines with high viscosity.
Clean the inhalation chamber a s described in section
“CARE, STORAGE, REPAIR AND RECYCLING”.
If after cleaning the problem persists, replace the
inhalation chamber to a new chamber.
Electrodes of nebulizer and
inhalation chamber are
contaminated.
Wipe and re-power on.
LED Watt blinks, charge level
of batteries in inadequate.
Replace all the batteries and re-power on
After power on,
indicator POWER
lights for one
second, and then
turned off
Inhalation chamber is empty
Pour the inhalation solution in to the inhalation
chamber.
No contact between inhalation
solution and mesh membrane.
Change the angle of inclination of nebulizer so
that the inhalation solution contacts with mesh
membrane.
Electrodes of nebulizer and
inhalation chamber are
contaminated.
Wipe and re-power on
Light POWER
is not on and
medicine is not
sprayed
Batteries are inserted
incorrectly
Insert the batteries observing polarity and again
turn on power.
Low battery charge
Replace all the batteries and re-power on