Chapter 6 Installation Instruction & Operating 

Instruction 

 
6.1 Installation Instruction 

1.

 

Determine the input voltage before using the power source. 

2.

 

Turn the voltage knob to the minimum voltage, i.e. rotate it counterclockwise to 
the minimum position. 

3.

 

Before connection to the power source, turn all switches to OFF to make sure 
the device has no possible damage. 

4.

 

The power source can only be connected when the device specification exactly 
match that of the power source system. 

 

6.2 Operating Instruction 

1.

 

Check whether control switches, knobs and displays are fine before using them. 

2.

 

Turn the power switch to the OFF position firstly. 

3.

 

Please determine again whether the input voltage is correct before connection 
to the power supply, to make sure the equipment is safe. 

4.

 

Internal AC contactor of the power supply picks up about 3 to 5 seconds after 
powering on. Press the Start button, and the output LED will light, indicating 
there is output voltage from power supply. Press again the Start button, and the 
output LED will be off, indicating there is no output voltage from the power 
supply. 

5.

 

Output Frequency Adjustment: you can turn the frequency selector to any 
position without powering on/off. The frequency indicated at fixed position is 
output frequency. To slightly adjust frequency, it is necessary to turn the fixed 
position to the rightmost and then adjust the spin button. 

6.

 

For example: To get the frequency of 52Hz, please firstly turn the selector to the 
VAR position, and then adjust the spin button until 52Hz appears on the display. 

7.

 

Nominal Voltage Adjustment: Turn the tri-position voltage selector to the Middle 
position, and then rotate the cross-type spin button on the right to get the 
desired voltage. 

8.

 

Connect the load to the terminal block, and press the Start button, and the 
output LED will light. In this way, you can easily get all values displayed on the 
functional display. 

9.

 

Notice (I): Over current, over load, over heat and short circuit protective 
devices are supplied with this machine. In case of over current, over load, over 
heat or short circuit, the protective circuit will immediately cut off the power and 
the buzzer will sound alarm. At this time, please firstly turn the load off, and 
then press the button Reset. The power supply begins providing the power after 
the alarm stops. Check the load conditions, and then turn the load on after 
determining there is no exception. 

10.

 

Notice (II): Warning! If it is necessary to make 220V/110V conversion (inside 

Summary of Contents for LSP-1KVA

Page 1: ...p com Lisun Electronics Shanghai Co Ltd Add 113 114 No 1 Building Nanxiang Zhidi Industry Park No 1101 Huyi Road Jiading District Shanghai 201802 China Tel 86 21 5108 3341 Fax 86 21 5108 3342 Email Sa...

Page 2: ...3 Block Diagram for Operation of LSP Series of Variable Frequency Power Supply 3 Chapter 4 Method for Selecting Power Capacity 3 Chapter 5 Instruction for Control Panel of Variable Frequency Power Sup...

Page 3: ...pply of required frequency The features of ideal AC power supply are stable frequency stable voltage zero internal resistance and Sine wave form without distortion Variable frequency power supply is c...

Page 4: ...ce Power Capacity 1 1 Load Power Load Starting Current 2 Induction Power Capacity Load Power Load Rated Current Wave Crest Coefficient for Load Current 3 Rectification Power Capacity Load Power 1 5 4...

Page 5: ...oltage between 10 and 30 less than the nominal voltage 10 Tri Position Voltage Selector High Voltage More Than Nominal Value Middle Nominal Voltage Low Voltage Less Than Nominal Value 11 Nominal Volta...

Page 6: ...requency Adjustment you can turn the frequency selector to any position without powering on off The frequency indicated at fixed position is output frequency To slightly adjust frequency it is necessa...

Page 7: ...e Phase Voltage Range HIGH 0 300V LOW 0 150V Output Frequency Adjustable between 45 0 70 0 Hz 50Hz 60Hz 100Hz 200Hz 400Hz Load Voltage Stabilization Rate 0 5 Frequency Stabilization Rate 0 10 Wave Dis...

Page 8: ...tment of our company Chapter 9 Precautions for Use The instruments should operate in the specified environments The installation environments need to be noticed considering the function and service li...

Page 9: ...com www Lisungroup com Page 8 Never power off at the instrument running Never contact the charged positions to avoid electric shock at the instrument running The personnel without professional trainin...

Reviews: