Microphone Preamplifier

SoundConnect comes standard with a Brüel & Kjær 7-pin LEMO 
connector. 

Table 1: Pin Connections for the Lemo Preamplifier Socket

Polarization Voltage

Just above the preamplifier input socket is a three-position slide switch for selecting 0, 28, or 200 volts 
microphone polarization. The 0 volt setting is for prepolarized condenser microphones.

Important!

The BNC input does not provide microphone powering, e.g.; IEPE. Microphone power is 
available only from the Lemo Connector.

Calibration

The ANSI S1.40 standard recommends that a certified calibration laboratory calibrate test instrumentation 
at least once a year. 

Service and Repair

SoundConnect is designed and constructed to provide the user with many years of safe and trouble free 
operation. However, should a fault occur which impairs its correct functioning and operating safety, it 
should be immediately shut off, disconnected at the mains source and disconnected from other hardware. 
For repair, contact your local LISTEN service representative. Under no circumstances should persons not 
qualified in the service of electronic instrumentation attempt repair. 

Pin

LEMO

1

Ground

2

Ground

3

Polarization voltage

4

Signal input

5

Not connected

6

Power supply positive

7

Power supply negative

Case

Connected to instrument chassis

 

6

5

7

1

2

3

4

 

Polarization 

Voltage

+14V DC

-14V DC

Figure  1-3:  Lemo Socket

Summary of Contents for SoundConnect

Page 1: ...ectTM Microphone Power Supply Instruction Manual April 2015 2006 2015 Listen Inc Sound Measurements Make Sound Products 580 Harrison Ave Suite 3W Boston MA 02118 617 556 4104 Fax 617 556 4145 www list...

Page 2: ...t to match the available mains voltage and that the correct fuse is installed Safety Symbols zThe apparatus will be marked with this symbol when it is important that the user refer to associated warni...

Page 3: ...andling of the Products against defects or damages arising from improper installation where installation is by persons other than Manufacturer against defects in products or components not manufacture...

Page 4: ...acquisition boards for converting the input signal into the digital domain Using SoundConnect s output attenuator and peak level meter it is easy to achieve the full performance from these boards Thr...

Page 5: ...peak 5 VRMS A3 should be set to the closest setting which best matches the data acquisition board s maximum input voltage specification If the board s limit falls in between A3 s settings select the l...

Page 6: ...oratory calibrate test instrumentation at least once a year Service and Repair SoundConnect is designed and constructed to provide the user with many years of safe and trouble free operation However s...

Page 7: ...tput must be connected to the 3 pin header at the top of the frontend pc board 2 Slide the case off of the base plate 5 Replace the case by sliding it into the first groove on each side of the base pl...

Page 8: ...Megohm z Output impedance 600 ohms z Output attenuator 0 dB 7 07 VP 8 dB 2 82 VP 14 dB 1 41 VP 0 1 dB z Peak Level Meter Indication at 20 10 0 dB and overload relative to selected output range 0 5 dB...

Reviews: