Technical Details
Factor
Target
Comments
Specificity
Specific to alcohol only
When used correctly, not
sensitive to acetone, paint
and glue evaporation, food,
sweets, methane and all
materials without any
alcohol content which can be
found in human breath
Measurement
range
0.02mg/L breath (BrAC)
Minimum
Note 1
2.00mg/L breath (BrAC)
Maximum
Measurement
Accuracy
±0.02mg/L
≤ 0.20mg/L
±10% above 0.20mg/L
In Active Mode only
Measurement
step
0.01mg/L
Response time
less than 60 seconds
0.22mg/L alcohol level
Recovery time
less than 100 seconds
0.22mg/L alcohol level
Sampling
frequency
Less than 25 seconds
Negative test result
Operating
temperature
range
-5
o
C
Minimum
40
o
C
Maximum