Linktop Otoscope NE20 User Manual Download Page 17

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic 
is affected by absorption and reflection from structures, objects and people.

Test 
Frequency 

(MHz)

385

450

710
745
780
810
870
930

380-

390

430-

470

704-

787

800-

960

1700-
1990

TETRA

400

GMRS  

460, 

FRS

460
LTE 13, 

17 

Band

GSM 

800/900

GSM 

800/900, 

TETRA  

800 , 

iDEN 820, 

CDMA

850, 

LTE 5 

Band
GSM 

1800 , 

CDMA 

1900, 

GSM 

1900, 

DECT,

LTE  1,  3,  

4  , 25

Band, 

UMTS

Pulse 

modulation 

18 Hz
FM ±5 KHz

deviation 

1KHz sine

Pulse 

modulation 

217 Hz

Pulse 

modulation 

217 Hz

Pulse 

modulation 

18 Hz

1.8

0.3

27

2

0.3

28

2

0.3

28

2

0.3

28

0.2

0.3

9

Band 

(MHz)

Service Modulation Maximum 

Power (W)

Distance
(m)

Immunity 
Test Level 
(V/m)

a   Field strengths from transmitters, such as base stations for radio 
(cellular/cordless) telephones and land mobile radios, amateur radio, AM 
and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF 
transmitters, an electromagnetic site survey should be considered. If the 
measured filed strength in the location in which the Otoscope is used 
exceeds the applicable RF compliance level above. The Otoscope should 
be observed to verify normal operation. If abnormal performance is 
observed, additional measures may be necessary, such as reorienting or 
relocating the Otoscope.

b    Over the frequency range 150kHz to 80 MHz, filed strengths should be 

less than 3V/m.

Recommended separation distances between portable and 

mobile

1720
1845
1970

Summary of Contents for Otoscope NE20

Page 1: ...Otoscope User Manual Version 1 0 Model NE20 ...

Page 2: ...re reference Linktop bears no liability and provides no guarantee with regard to damage including indirect damage arising due to this user manual not being complied with Linktop reserves the right to make technical changes without notice in the interest of progress Prior notices will not be given in case of any amendments within this manual ...

Page 3: ...Information 1 2 Description 3 3 Using Instructions 6 4 Cleaning and Maintenance 8 5 Environment Conditions 8 6 Troubleshooting 9 7 Servicing 9 8 Warranty 9 9 Explanation of utilized symbols 10 10 Electromagnetic Compatibility 11 ...

Page 4: ...tute a strangulation hazard any examination of children should be conducted under supervision Use only original Otoscope parts spare parts and accessories The correct and safe function of the device is only guaranteed if both the device used and its accessories are provided by Linktop Do not perform examinations with the device or put it to the body while it is being charged This medical product i...

Page 5: ...liable which may lead to their content being interpreted incorrectly by a doctor and in consequence to incorrect diagnosis The list of approved applications is available at the www linktop com website In case of noticing any irregularities in functioning of the device or any damage to it stop using it immediately and contact the manufacturer Do not touch the lens directly by hand or something hard...

Page 6: ...era which makes the light path and increase the visual field of view It has rechargeable battery powered It needs to be used with the USB cables to connect to Android system equipment and iPhone By using visualization App it enables capturing and transmission of video and still images It can be used on any person undergoing a physical examination It is suitable for adulthood and children over 3 ye...

Page 7: ... a thin rod into the reset hole to short press the reset key Indicator light Two color LED light used to indicate the operation charging and low power status of the device Magnetic USB interface For connecting charging cable Android iOS device Green led is on when the device is on and connected to an iOS device 1 2 3 4 5 6 Green led flashes slowly when the battery is below 20 low power mode and no...

Page 8: ... 1A 500mA 24 5 g Battery life Diameter 3 9mm Pixel 1 000 000 pixels Resolution 720p Focal length 1 5 2 0cm Horizontal viewing angle FOV 70 Frame rate 15fps Otoscope 1 User manual 1 Magnetic USB C cable 1 Magnetic iOS cable 1 Micro USB adapter cable 1 Magnetic USB charging cable 1 Height 5 3 inches 134 5mm Diameter 0 7 inches 17 8 mm Otoscope NE20 Lithium ion battery non removable ...

Page 9: ...nload the application from Google Play or App Store Please make sure that the app permissions of the mobile phone is open Attention Use only the enclosed charging cable In any other case the charging cable should be connected to a charger that meets the requirements described in this manual When charging is complete please unplug the charging cable 3 1 2 Charge the battery When using the Otoscope ...

Page 10: ... C cable 3 Once you have connected the Otoscope to your mobile phone open the application the application will automatically bind the device 4 Use the application to start to preview captured images 5 Point the camera close to the specific spots you want to observe Figure 4 and hold the Otoscope at each spot use the application to take photos or shoot videos until you have finalized the observatio...

Page 11: ...eric pressure 700 hPa 1060 hPa 5 2 Operation conditions Temperature 5 to 35 Relative humidity 10 to 95 Atmospheric pressure 700 hPa 1060 hPa 3 3 Turning the Otoscope off If you do not use the Otoscope you can remove the cable and close the cover to turn the Otoscope off Attention If the Otoscope is not turned off when it is not in use it will run out of battery power The Otoscope should be cleaned...

Page 12: ... in functioning of the medical product contact the maintenance service at the e mail address support linktop com Linktop bears no liability and provides no guarantee with regard to damage including indirect damage arising due to this user manual not being complied with 8 Warranty The warranty for the entire product is invalidated if non genuine Linktop products or non original parts are used and i...

Page 13: ... to instruction manual booklet Caution IP classification Type BF Applied Part Medical device Part recycled separately from other waste Do not use if package is damaged Fragile handle with care Humidity limitation Atmospheric pressure limitation Temperature limit Keep away from sunlight Keep dry Direct current Authorized representative in the European Community European Union 6 7 8 9 11 12 13 14 15...

Page 14: ...ration in the configuration in which it will be used The Otoscope is intended for use in the electromagnetic environments specified below The customer or the user of the Otoscope should assure that it is used in such an environment Emissions test RF emissions CISPR 11 RF emissions CISPR 11 Group 1 Class B Harmonic emissions IEC 61000 3 2 Voltage function flicker emissions IEC 61000 3 3 Not applica...

Page 15: ...dip in UT for 25 cycles 5 UT 95 d i p in UT for 5 sec Not applicable Battery operated device Not applicable Battery operated device Not applicable Battery operated device Mains power quality should be that of a typical commercial or hospital environment Mains power quality should be that of a typical commercial or hospital environment Mains power quality should be that of a typical commercial or h...

Page 16: ... used no closer to any part of the Otoscope including cables than the recommended separation distance calculated from the equation application to the frequency of the transmitter Recommended separation distance Where p is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in meter m Field strengths fo...

Page 17: ... Maximum Power W Distance m Immunity Test Level V m a Field strengths from transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered ...

Page 18: ...Ltd Address Room 501 2 502 503 North Building Torch Hi Tech Zone No 56 58 Huoju Road Xiamen 361000 Fujian P R China Website www linktop com 2 0 3 28 2 0 3 9 Bluetooth WLAN 802 1 1 b g n RFID 2450 LTE 7 Band WLAN 802 11 a n Pulse modulation 217 Hz Pulse modulation 217 Hz ...

Page 19: ......

Reviews: