LIV® IQ Linde Integrated Valve with digital display
– portable, delivery system for Medicinal Oxygen.

Linde: Living healthcare

Summary of Contents for LIV IQ

Page 1: ...Manual LIV IQ Linde Integrated Valve with digital display portable delivery system for Medicinal Oxygen Linde Living healthcare...

Page 2: ...02 Manual LIV IQ Oxygen This manual was last revised July 2015 LIV is a registered trademark of The Linde Group...

Page 3: ...uction 04 2 Device Description 04 3 Intended Use 05 4 Safety Requirements 06 5 Operating Instructions 08 6 MR Compatibility 08 7 Cleaning 08 8 Service and Maintenance 09 9 Fault Cause Remedy 11 10 Sto...

Page 4: ...read this instruction carefully before using the LIV IQ for the first time 3 Intended Use The LIV IQ is suitable for use in all healthcare settings including but not limited to hospital out patient im...

Page 5: ...dropped ADDITIONAL INFORMATION Do not move the LIV IQ package during MRI scan as the MR image quality can be degraded If you suspect there may be electro magnetic interference between the LIV IQ and o...

Page 6: ...as engaged and is not in between two settings Otherwise the flow regula tor will not allow the required volumetric flow rate for the medicinal gas 3 Flow selector 4 Flow outlet 2 Shut off valve 1 Pres...

Page 7: ...dition occurs Set ow selector to zero before opening shut off valve Warning triangle G H A B C D F E G H I J K L M O N Full full full full New full cylinder diamond Content indicator B A As gas is con...

Page 8: ...cleaned the cleaning solutions used must be compatible with brass plastic materials of components including labels and oxygen Do not use cleaning solutions containing ammonia Do not immerse the produc...

Page 9: ...2 Make sure that the flow selector is set at zero close the shut off valve and mark the cylinder as defective Put the product in a well ventilated area and contact Linde Healthcare Battery is low Use...

Page 10: ...product in a well ventilated area and contact Linde Healthcare alternating in the flow rate area of the display An audible signal sounds 1 Two devices draw gas simultaneously 2 One device draws gas 1...

Page 11: ...medicinal gas after its expiry date see product label Warning notices prohibiting smoking and naked lights must be posted clearly in the cylinder storage area and the Emergency Services should be adv...

Page 12: ...Linde Gas North America LLC 575 Mountain Ave Murray Hill NJ 07974 Phone 866 543 3427 www linde healthcare us MED 769en US 2015 09...