5. Operating Instructions.

Pressure outlet, 1) – additional connection information

If provided, the pressure outlet is equipped with a quick-connector and
allows connection to a medical device. Output pressure 50 psi nominal. The
flow selector does not control the flow from the pressure outlet.

To connect a delivery device to the quick-connector (pressure outlet):

Make sure that the probe connector is compatible with the respective
medicinal gas.

Insert the probe connector into the quick-connector so that it engages
correctly. Pull on the connector to make sure that it is sitting correctly.

This quick-connector does not have a function to retain the inserted probe.
Therefore, do not disconnect the connector until the valve is closed and
depressurized.

A faulty connection can result in leaks and/or to the ejec-
tion of the connector, with potential injuries/damage.

Flow outlet, 4) – additional connection information

To connect a delivery device to the flow outlet, make sure that:

The patient is not connected to the delivery device.

The dimension of the delivery device hose fits the flow outlet of the
valve (6 mm outer diameter).

The delivery device hose is firmly attached. 
Push delivery device hose onto the flow outlet of the LIV IQ and make
sure that it is firmly attached before opening the shut-off valve.

The flow selector dial has engaged.
An audible alert and the words “Set FLO” will appear on the display if
the flow selector is between settings.

Make sure that the flow selector dial has engaged and is
not in between two settings. Otherwise, the flow regula-
tor will not allow the required volumetric flow rate for
the medicinal gas.

3) Flow selector

4) Flow outlet

2) Shut-off valve

1) Pressure outlet equipped with

quick-connector (if provided)

General instructions for use

All personnel handling cylinders with integrated valves should
be fully trained to carry out each operation and be fully aware
of the requirements detailed in this  instruction for use.

Operating temperature –4 °F to 113 °F (–20 °C to +45 °C).

Only use with devices intended for the product administration.

Prepare to change the cylinder when the digital display contents
indicator shows only the segments in the red area.  

To ensure the correct, prescribed flow rate, always make sure that the
flow selector is positioned on a numbered flow setting, not between
flow settings.

5) Digital display

6) Mute button

Before Use

Make sure that the digital display (5) contents indicator shows there is
enough content for intended use.

Make sure that the flow selector (3) is set at zero.

Connect the delivery device to outlet (1 or 4), see below.

Slowly turn the shut-off valve (2) anticlockwise to open the LIV IQ fully.

If a delivery device is connected to the flow outlet (4), turn flow
 selector (3) stepwise up through flow settings to select flow.

The patient can now be connected to the delivery device.

After Use

At conclusion of treatment or at exchange of an empty cylinder:

Remove the delivery device from the patient.

Turn flow selector (3) stepwise down through flow settings to zero.

Turn the shut-off valve (2) clockwise to close the LIV IQ.

Turn the flow selector (3) fully to vent the equipment.

Turn the flow selector (3) back to zero.

Disconnect the delivery device from outlet (1 or 4 ).

Summary of Contents for LIV IQ

Page 1: ...Manual LIV IQ Linde Integrated Valve with digital display portable delivery system for Medicinal Oxygen Linde Living healthcare...

Page 2: ...02 Manual LIV IQ Oxygen This manual was last revised July 2015 LIV is a registered trademark of The Linde Group...

Page 3: ...uction 04 2 Device Description 04 3 Intended Use 05 4 Safety Requirements 06 5 Operating Instructions 08 6 MR Compatibility 08 7 Cleaning 08 8 Service and Maintenance 09 9 Fault Cause Remedy 11 10 Sto...

Page 4: ...read this instruction carefully before using the LIV IQ for the first time 3 Intended Use The LIV IQ is suitable for use in all healthcare settings including but not limited to hospital out patient im...

Page 5: ...dropped ADDITIONAL INFORMATION Do not move the LIV IQ package during MRI scan as the MR image quality can be degraded If you suspect there may be electro magnetic interference between the LIV IQ and o...

Page 6: ...as engaged and is not in between two settings Otherwise the flow regula tor will not allow the required volumetric flow rate for the medicinal gas 3 Flow selector 4 Flow outlet 2 Shut off valve 1 Pres...

Page 7: ...dition occurs Set ow selector to zero before opening shut off valve Warning triangle G H A B C D F E G H I J K L M O N Full full full full New full cylinder diamond Content indicator B A As gas is con...

Page 8: ...cleaned the cleaning solutions used must be compatible with brass plastic materials of components including labels and oxygen Do not use cleaning solutions containing ammonia Do not immerse the produc...

Page 9: ...2 Make sure that the flow selector is set at zero close the shut off valve and mark the cylinder as defective Put the product in a well ventilated area and contact Linde Healthcare Battery is low Use...

Page 10: ...product in a well ventilated area and contact Linde Healthcare alternating in the flow rate area of the display An audible signal sounds 1 Two devices draw gas simultaneously 2 One device draws gas 1...

Page 11: ...medicinal gas after its expiry date see product label Warning notices prohibiting smoking and naked lights must be posted clearly in the cylinder storage area and the Emergency Services should be adv...

Page 12: ...Linde Gas North America LLC 575 Mountain Ave Murray Hill NJ 07974 Phone 866 543 3427 www linde healthcare us MED 769en US 2015 09...