Portable concentrators. Equipment guide.
Portable concentrators. Equipment guide.
Mode icons – top left corner of display
One of two icons can be displayed in the
top left corner of the Inogen One
®
G3’s
display window. Both relate to whether
or not the breath detection audible alarm
is enabled or not.
Inogen One
®
G3 system notifications
The Inogen One
®
G3 monitors various
parameters during operation and utilises
an intelligent alarm system to indicate
concentrator malfunction. If multiple alarm
conditions are detected, the highest priority
alarm will be displayed. The following
notification messages are accompanied by
a single, short beep:
Inogen One
®
G3 low priority alerts
The Inogen device has three types of priority
alart: low, medium and high. Low priority
alerts typically mean the device can still be
used but may require attention. Medium and
high priority alerts require immediate contact
with BOC Homecare on
0800 136 603
.
The information on the following pages
provides more detail. In all cases, please
contact BOC Homecare on
0800 136 603
to
discuss any questions you may have.
Low priority alert messages (see table below)
are accompanied by a double beep with a
solid red light showing on the display panel.
*intake (blue plastic surround)
can vary depending on
device iteration
Icon
Meaning
22
Power Status Icons (continued)
The icons below are examples of those shown when the Inogen One® G3 is operating
from an external power supply and charging the battery. The lightning bolt indicates
that an external power supply is connected.
!
C
!
C
!
C
Meaning
Battery is charging with charge level between 60% and 70%.
The battery is fully charged and is charging as necessary to
maintain its charge.
Battery is charging with charge level less than 10%.
The Inogen One® G3 is operating from an external power source with
no battery present.
Icon
Mode Icons
These are the icons shown in the display’s mode window.
Display Text
NOTE
When two conditions occur simultaneously, the condition with the
higher priority will be displayed.
Meaning
The breath detection audible alarm has been enabled.
The breath detection audible alarm is disabled. This is the default condition.
Icon
!
C
The breath detection
audible alarm has been
enabled.
22
Power Status Icons (continued)
The icons below are examples of those shown when the Inogen One® G3 is operating
from an external power supply and charging the battery. The lightning bolt indicates
that an external power supply is connected.
!
C
!
C
!
C
Meaning
Battery is charging with charge level between 60% and 70%.
The battery is fully charged and is charging as necessary to
maintain its charge.
Battery is charging with charge level less than 10%.
The Inogen One® G3 is operating from an external power source with
no battery present.
Icon
Mode Icons
These are the icons shown in the display’s mode window.
Display Text
NOTE
When two conditions occur simultaneously, the condition with the
higher priority will be displayed.
Meaning
The breath detection audible alarm has been enabled.
The breath detection audible alarm is disabled. This is the default condition.
Icon
!
C
The breath detection
audible alarm is disabled.
This is the default
condition.
Message
display and
text
Meaning
Inogen One
®
G3 shutting
down
ON/OFF button has been
pressed for two seconds.
Concentrator is performing
system shut down.
HH:MM Vx.x:
Serial Number
The alarm button has been
pressed for five seconds.
Message display and text
Condition/action/explanation
Check battery
Battery error has occured. Check the connection of your battery
and ensure that it is properly attached and latched onto the
concentrator. If battery error recurs with the same battery, stop
using the battery and switch to a new battery or remove and
operate concentrator using external power supply.
Oxygen low see manual
Concentrator is producing oxygen at a slightly low level (<82%)
for a period for 10 minutes. If condition persists, contact your
equipment provider.
Remove battery to cool
Battery has exceeded its charging temperature and charging has
stopped. The battery will not charge while this alert is present
but will begin to charge when the battery temperature returns
to the normal operating range. If the battery charging is desired
sooner, remove the battery from the concentrator and allow it
to cool in an open area for approximately 10-15 minutes. Then,
re-insert the battery into the Inogen One
®
G3. If the problem
still persists, contact BOC Homecare on
0800 136 603
.
Comm error see manual
Concentrator is producing oxygen but cannot report battery
status. Replace battery. If condition persists, contact BOC
Homecare on
0800 136 603
.
Service needed
The concentrator requires servicing at the earliest convenience.
The concentrator is operating to specification and may continue
to be used. Contact BOC Homecare to arrange service.
o
2
sensor fail see manual
The concentrator’s oxygen sensor has malfunctioned. You may
continue to use the concentrator. If the condition persists,
contact BOC Homecare on
0800 136 603
.