Lincoln Electric ARC TRACKER K3019-1 Operator'S Manual Download Page 13

 

English  

English 

13

 

Adjustment Procedure: 

 

1.  Volt meter Display 
2.  Volt Meter Calibration Adjust 
3. Calibration 
4. Right 

Push 

Button 

5. Center 

Knob 

6. Left 

Push 

Button 

7.  Ammeter Calibration Adjust 
8. Ammeter 

Display 

Figure D.2: Front Panel 

 

A.  With the power source off, connect the ARC 

TRACKER™ being calibrated to input power 
and turn the ARC TRACKER™ on. 

B.  Enter into the calibration menu by pressing 

both the left and right push buttons on the ARC 
TRACKER ™ user-interface. The Calibration 
Mode LED will illuminate and the message 
display will look like this: 

 

C.  Turn the center knob until parameter P.83 is 

reached. Press the right pushbutton to enter 
into Calibration mode. 

 

 

D.  With the power source turned off, press the 

right pushbutton to zero the unit. The unit will 
take a few seconds to zero. 

 

 

 

E.  Turn on the power source to provide the 

maximum calibration current and voltage, e.g. 
1000 A, 44 V. The ARC TRACKER™ will 
begin to display the voltage, amperage, & arc 
time, and will begin to calculate and display 
heat input. 

F.  6. Press the right push button to begin 

calibration of the volt meter and ammeter 
located on the ARC TRACKER™ under 
calibration. 

 

 

G.  7. Turn the Volt Meter Calibration Adjust knob 

located under the volt meter on the user-
interface until it matches the reference volt 
meter. 

H.  Turn the Ammeter Calibration Adjust knob 

under the ammeter on the user-interface until it 
matches the reference ammeter. 

I.  When both the volts display and amps display 

on the unit being calibrated are dialed in, press 
the right pushbutton to accept the calibration 
values. 

J.  Turn the center knob until parameter P.0 is 

displayed. Press the left push button to EXIT. 

K.  11. Verify that the calibration was completed 

successfully by adjusting the power source to 
different set points for current and voltage 
comparing the volts display and amps display 
on the unit being calibrated to the reference 
volt meter and ammeter. See Table D.1. 

L.  12. As left values should be entered into Table 

D.1. Then the % deviation can be calculated 
by comparing the reference meter values to 
the ARC TRACKER™ as left values. 

Summary of Contents for ARC TRACKER K3019-1

Page 1: ...English English 1 IM3047 06 2017 REV01 ARC TRACKERTM OPERATOR S MANUAL ENGLISH THE LINCOLN ELECTRIC COMPANY 22801 St Clair Ave Cleveland Ohio 44117 1199 USA www lincolnelectric eu...

Page 2: ...ference record in the table below your equipment identification information Model Name Code Serial Number can be found on the machine rating plate Model Name Code Serial number Date Where Purchased EN...

Page 3: ...OO ST STO STOO OR EQUIVALENT EXTRA HARD USAGE CORD NEMA 5 15P INCLUDED 230 0 5A ANY NEMA 250V TYPE 230 0 5A EUROPE 3 CONDUCTOR 1 0 mm2 HAR CEE 7 7 All attachment plugs must comply with the Standard fo...

Page 4: ...K3019 number may also contain prefixes and suffixes Is in conformity with Council Directives and amendments Electromagnetic Compatibility EMC Directive 2014 30 EU Low Voltage Directive 2014 35 EU Stan...

Page 5: ...surement Personal medical devices like pacemakers and hearing aids Check the electromagnetic immunity for equipment operating in or near the work area The operator must be sure that all equipment in t...

Page 6: ...AGNETIC FIELDS MAY BE DANGEROUS Electric current flowing through any conductor creates electric and magnetic fields EMF EMF fields may interfere with some pacemakers and welders having a pacemaker sha...

Page 7: ...itable Location UNIT IS IP23 RATED The ARC TRACKER will operate in harsh environments Even so it is important that simple preventative measures are followed in order to assure long life and reliable o...

Page 8: ...he power source and the ARC TRACKER right side weld terminals when viewed from the rear Connect cable s of sufficient size and length from the ARC TRACKER left weld terminals to the work Be sure the c...

Page 9: ...e ARC TRACKER has been calibrated before being shipped from the factory The Lincoln Electric Company recommends that end users of its welding equipment evaluate the suitability of utilizing this produ...

Page 10: ...press the Right hand button to save the new setting Pressing the Left button will cancel the change C To exit the set up menu at any time press the Right and Left buttons of the Main Display panel sim...

Page 11: ...s If other meters are used the compatibility and accuracy with inverter power sources must be determined by the user Volt meter Keithley 2701 Digital Multimeter Ammeter Keithley 2701 Digital Multimete...

Page 12: ...ter Value 100 Dev As Left ARC TRACKER As Left ARC TRACKER As Found Reference Meter Value 100 Table D 1 Nominal Set Point Reference Meter Value Arc Tracker As Found Deviation As Found Arc Tracker As Le...

Page 13: ...erage arc time and will begin to calculate and display heat input F 6 Press the right push button to begin calibration of the volt meter and ammeter located on the ARC TRACKER under calibration G 7 Tu...

Page 14: ...r any customer s particular purpose with respect to such information or advice As a matter of practical consideration we also cannot assume any respon sibility for updating or correcting any such info...

Page 15: ...et information on approved collection systems from our local representative By applying this European Directive you will protect the environment and human health Spare Parts 12 05 For Spare Parts refe...

Page 16: ...ence only It may not be accurate for all machines covered by this manual The specific diagram for a partiuclar code is pasted inside the machine on one of the enclosure panels If the diagram is ellegi...

Reviews: