Page 15 of 31
Mounting the TD5 Standard
Mounting is made by pushing the TD5 unit into the bed bracket and thereafter slide the closing devices on.
It is IMPORTANT to be completely sure that the closing devices are in place before using the TD5.
Recommendations
• Mains cable and controls must be ordered separately
• The shaft diameter must be Ø25.2 (+0.3/-0.2) mm
• Always use the cable relief for the mains cable on the TD5 to avoid pulling out the mains cable plug from
the TD5 resulting in a malfunction.
• Do not look directly into the light under the bed.
• The default timeout for the Under Bed Light is 30 minutes.
• Do not drive the TD5 system without the closing devices mounted
• The bed brackets must have a smooth surface without burs and welding blops in the area where the TD5 is mounted.
• It is important to turn the key in the direction indicated on the housing. Turning the key in the wrong direction can
wreck the inner hex and make mechanical emergency lowering impossible.
• Do not use an electric screwdriver for lowering, use the enclosed hex key on the TD5.
• The TD5 must be in fully retracted position when starting to lifetime simulator.
1. Push TD5 unit into bed bracket.
2. Slide each closing device on and press it into correct position.
IMPORTANT:
Ensure that the closing devices are in place before using the TD5.
Do not drive the TD5 system without the closing devices mounted.
The bed brackets must have a smooth surface without burs and welding blops in the area where the TD5 is mounted.