48
Alarm stop
In case certain important sterilization parameters are not met, the sterilizer will generate an alarm code
and abort the cycle automatically.
The sterilizer enters into a reset phase; a wait message and an alarm code are displayed on the screen.
Do not switch off the sterilizer: It will take some time (several minutes) to reset the system and reach safe conditions in the sterilizer
chamber before it is possible to open the sterilizer door and remove the load.
Alarm end
When the reset phase is over, you will be asked to press BACK to get to the
“Unlock door” option.
Confirm (OK) to unlock the door.
While the door is unlocking, a waiting message is displayed.
Water could be present in the chamber when
opening the door: prevent spilling (e.g.,
place a towel under the chamber door).
At this stage select and confirm “Info” to view the sterilizer parameters (see Chapter 6 of this manual).
Open the chamber door
and remove the load.
The message LOAD NOT STERILE means that the load is not sterile. Do not use
items on patients!
The message DRYING INTERRUPTED means that the load might be wet. Wet items
are for immediate use only!